Whole-Brain SNR-Efficient Pseudo-Continuous Arterial Spin Labeling at 7T
Motivation: The high SAR burden of PCASL at 7T typically requires long TRs and short label durations, reducing SNR efficiency and perfusion measurement accuracy. Goal(s): Our goal was to maximize the SNR efficiency of PCASL at 7T. Approach: We optimized the PCASL B1+/gradient parameters and background suppression pulses to maximize SNR efficiency (i.e., balancing labeling efficiency and SAR), and utilized VERSE and dynamic B0 shimming to reduce SAR. Results: Compared to optimizing the PCASL parameters for maximum labeling efficiency, our max-SNR efficiency optimized parameters increased SNR efficiency by 38% and reduced TRs by 41% in vivo. Impact: The improved SNR efficiency of our optimized 7T PCASL parameters increases the advantages of ultra-high field perfusion measurements, enabling robust and high-SNR perfusion measurement in clinically viable scan times.
Authors
- Aaron T. Hess (ORCID: https://orcid.org/0000-0002-9289-5619)
- Joseph G. Woods (ORCID: https://orcid.org/0000-0002-0329-824X)
- Yang Ji (ORCID: https://orcid.org/0000-0003-4134-374X)
- LI Hong-we (ORCID: https://orcid.org/0000-0003-1119-2119)
- Thomas W. Okell (ORCID: https://orcid.org/0000-0001-8258-0659)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/2013
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00