Up to 24-channel receiver coil arrays for concurrent TMS-MRI imaging at 3T
Motivation: Concurrent TMS-fMRI lacks whole-brain coverage or has low SNR. Goal(s): Develop setups allowing for whole-brain TMS-fMRI with high spatiotemporal resolution and minimizing the interference between MRI and TMS during data collection and neuromodulation. Approach: Combinations of 8- and 16-channel arrays enabled whole-brain TMS-MRI with 16 or 24 channels. Results: Structural and functional images of high spatiotemporal resolution (1.5 mm with 2 s/volume or 5 mm with 0.1 s/volume) were demonstrated with an MRI-compatible TMS coil. Impact: We developed concurrent TMS-MRI setups with whole-brain coverage and high spatiotemporal resolution using combinations of 8- and 16-channel coil arrays. Images of high spatiotemporal resolution were demonstrated.
Authors
- Fa‐Hsuan Lin (ORCID: https://orcid.org/0000-0002-9539-1731)
- Hsin-Ju Lee (ORCID: https://orcid.org/0000-0003-3358-8720)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/2722
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00