Improved visualization of high b-value prostate diffusion weighted imaging using LIPO-only multi-shot EPI with 2D navigator (LION-IRIS)
Motivation: One of the drawbacks of IRIS is its low SNR compared with ss-DWI. High b-value DWI improves the ability of prostate cancer detection by increased cancer conspicuity. However, such high b value images may also suffer from decreased SNR. Goal(s): We hypothesized that using only slice selection gradient reversal (SSGR, LIPO) technique without fat suppression pre-pulse (LIpo-ONly :LION) could improve the SNR of high b-value IRIS DWI. Approach: SNR,CR and ADC values were calculated with ROIs placed in internal obturator muscle and prostate region in peripheral zone. Results: LION IRIS-DWI could provide high quality distortion-free high b-value prostate DWI sufficient SNR. Impact: The present findings suggest that LION IRISs-DWI sequence demonstrated improved image distortion and blurring compared to ss-DWI, with having sufficient SNR and CR.It may help to further assess the prostate cancer pathology.
Authors
- Masami Yoneyama
- Iain Ball
- Daichi Murayama
- Takayuki Sakai (ORCID: https://orcid.org/0000-0002-8170-8320)
- Shigehiro Ochi
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/5107
- Primary Topic
- MRI in cancer diagnosis
- Type
- article
- Field-Weighted Citation Impact
- 0.00