Open-window MSR design with active magnetic compensation coil based on COMSOL multiphysics

Magnetic shielding room (MSR) is one of the necessary equipment for measuring magnetocardiography (MCG) and magnetoencephalography (MEG). The traditional closed MSR only focuses on the indicators of the shielded magnetic field, and ignores the patient's feelings within the MSR. In this paper, an open-window MSR is proposed to solve the problem that patients have difficulty in communicating with the outside of closed MSR and have closed fear. The simulation results show that the size and distribution trend of the residual magnetic field in the inner working area of the MSR with shield-duct and active magnetic compensation coil is similar to that of the closed MSR. The method proposed in this paper provides a way to improve the comfort of patients in MSR for measuring MCG and MEG in humans.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2024-11-26
DOI
https://doi.org/10.1088/2057-1976/ad974a
Citations
1
Primary Topic
Atomic and Subatomic Physics Research
Type
article
Field-Weighted Citation Impact
0.34

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Open-window MSR design with active magnetic compensation coil based on COMSOL multiphysics

Peiling Cui, Zhouqiang Yang, Yanbin Li
1 citations
Biomedical Physics & Engineering Express
Atomic and Subatomic Physics Research
0.34
article

Open-window MSR design with active magnetic compensation coil based on COMSOL multiphysics

Peiling Cui, Zhouqiang Yang, Yanbin Li
article en
1 citations

Abstract

Magnetic shielding room (MSR) is one of the necessary equipment for measuring magnetocardiography (MCG) and magnetoencephalography (MEG). The traditional closed MSR only focuses on the indicators of the shielded magnetic field, and ignores the patient's feelings within the MSR. In this paper, an open-window MSR is proposed to solve the problem that patients have difficulty in communicating with the outside of closed MSR and have closed fear. The simulation results show that the size and distribution trend of the residual magnetic field in the inner working area of the MSR with shield-duct and active magnetic compensation coil is similar to that of the closed MSR. The method proposed in this paper provides a way to improve the comfort of patients in MSR for measuring MCG and MEG in humans.

Biomedical Physics & Engineering ExpressVol. 11(1)
Beihang University (CN)
National Natural Science Foundation of China
Openalex Percentile: Top 38%
Atomic and Subatomic Physics Research
0.34
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.