Combined drug delivery and treatment monitoring using a single high frequency ultrasound system

Ultrasound-mediated drug delivery is typically performed using transducers with center frequencies <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:mrow><mml:mo>≤</mml:mo></mml:mrow></mml:math>1 MHz to promote acoustic cavitation. Such frequencies are not commonly used for diagnostic ultrasound due to limited spatial resolution. Therefore, delivery and monitoring of therapeutic ultrasound typically requires two transducers to enable both treatment and imaging. This study investigates the feasibility of using a single commercial ultrasound imaging transducer operating at 5 MHz for both drug delivery and real-time imaging. We compared a single-transducer system (STS) at 5 MHz with a conventional dual-transducer system (DTS) using a 1.1 MHz therapeutic transducer and an imaging probe. <i>in vitro</i> experiments demonstrated that the STS could achieve comparable extravasation depth and area as the DTS, with higher drug deposition observed at 5 MHz. Additionally, extravasation patterns were influenced by peak negative pressure (PNP) and duty cycle, with the narrower beam width at 5 MHz offering potential advantages for targeted drug delivery. <i>in vivo</i> experiments in a murine bladder cancer model confirmed the efficacy of the STS for real-time imaging and drug delivery, with cavitation dose correlating with drug deposition. The results suggest that a single-transducer approach may enhance the precision and efficiency of ultrasound-mediated drug delivery, potentially reducing system complexity and cost.

Authors

Institutions

Publication Details

Journal
International Journal of Hyperthermia
Published
2024-11-26
DOI
https://doi.org/10.1080/02656736.2024.2430330
Citations
2
Primary Topic
Ultrasound and Hyperthermia Applications
Type
article
Field-Weighted Citation Impact
0.37

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Combined drug delivery and treatment monitoring using a single high frequency ultrasound system

Michael Gray, Christophoros Mannaris, Jia-Ling Ruan, Anne E. Kiltie et al.
2 citations
International Journal of Hyperthermia
Ultrasound and Hyperthermia Applications
0.37
article

Combined drug delivery and treatment monitoring using a single high frequency ultrasound system

Michael Gray, Christophoros Mannaris, Jia-Ling Ruan, Anne E. Kiltie, Richard J. Browning, Borivoj Vojnovic, Luca Baù, Eleanor Stride, Yesna O. Yildiz
article en
2 citations

Abstract

Ultrasound-mediated drug delivery is typically performed using transducers with center frequencies <mml:math xmlns:mml="http://www.w3.org/1998/Math/MathML"><mml:mrow><mml:mo>≤</mml:mo></mml:mrow></mml:math>1 MHz to promote acoustic cavitation. Such frequencies are not commonly used for diagnostic ultrasound due to limited spatial resolution. Therefore, delivery and monitoring of therapeutic ultrasound typically requires two transducers to enable both treatment and imaging. This study investigates the feasibility of using a single commercial ultrasound imaging transducer operating at 5 MHz for both drug delivery and real-time imaging. We compared a single-transducer system (STS) at 5 MHz with a conventional dual-transducer system (DTS) using a 1.1 MHz therapeutic transducer and an imaging probe. <i>in vitro</i> experiments demonstrated that the STS could achieve comparable extravasation depth and area as the DTS, with higher drug deposition observed at 5 MHz. Additionally, extravasation patterns were influenced by peak negative pressure (PNP) and duty cycle, with the narrower beam width at 5 MHz offering potential advantages for targeted drug delivery. <i>in vivo</i> experiments in a murine bladder cancer model confirmed the efficacy of the STS for real-time imaging and drug delivery, with cavitation dose correlating with drug deposition. The results suggest that a single-transducer approach may enhance the precision and efficiency of ultrasound-mediated drug delivery, potentially reducing system complexity and cost.

International Journal of HyperthermiaVol. 41(1)
University of Oxford (GB), Oxford BioMedica (United Kingdom) (GB)
Engineering and Physical Sciences Research Council
Good health and well-being
Openalex Percentile: Top 42%
Ultrasound and Hyperthermia Applications
0.37
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.