UV Spectrophotometric Method Development and Validation for the Determination of Nirmatrelvir Antiviral Drug in Bulk and its In-house Formulation

UV spectrophotometry was performed by Lab India Analytical UV 3092 using methanol as medium. The resolution process was performed in the range of 200–400nm to determine the maximum absorption (λ max). The λmax of Nirmatrelvir was found to be 210nm. Measure the absorbance of 10, 20, 30, 40, 50, and 60µg/ml drug solution and measure the maximum absorbance at 210 nm with a blank control. Then draw a graph from the concentration on the X-axis and the absorbance on the Y-axis to get a straight line. Validation parameters such as linearity, precision, specificity, LOD and LOQ, accuracy and robustness were evaluated according to ICH guidelines.

Authors

Institutions

Publication Details

Journal
Asian Journal of Pharmaceutical Analysis
Published
2024-09-20
DOI
https://doi.org/10.52711/2231-5675.2024.00022
Citations
1
Primary Topic
Click Chemistry and Applications
Type
article
Field-Weighted Citation Impact
0.25
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

UV Spectrophotometric Method Development and Validation for the Determination of Nirmatrelvir Antiviral Drug in Bulk and its In-house Formulation

Vikram Jadhav, Ashish Jain, Vaishali Jadhav, Priya R. Wani
1 citations
Asian Journal of Pharmaceutical Analysis
Click Chemistry and Applications
0.25
article

UV Spectrophotometric Method Development and Validation for the Determination of Nirmatrelvir Antiviral Drug in Bulk and its In-house Formulation

Vikram Jadhav, Ashish Jain, Vaishali Jadhav, Priya R. Wani
article en
1 citations

Abstract

UV spectrophotometry was performed by Lab India Analytical UV 3092 using methanol as medium. The resolution process was performed in the range of 200–400nm to determine the maximum absorption (λ max). The λmax of Nirmatrelvir was found to be 210nm. Measure the absorbance of 10, 20, 30, 40, 50, and 60µg/ml drug solution and measure the maximum absorbance at 210 nm with a blank control. Then draw a graph from the concentration on the X-axis and the absorbance on the Y-axis to get a straight line. Validation parameters such as linearity, precision, specificity, LOD and LOQ, accuracy and robustness were evaluated according to ICH guidelines.

Asian Journal of Pharmaceutical Analysis
Central Drug Research Institute (IN), Advanced Device Technology (United States) (US), Pharmaceutical Product Development (India) (IN)
Good health and well-being
Openalex Percentile: Top 46%
Click Chemistry and Applications
0.25
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.