Aortic strain, flow pattern and wall shear stress in a patient-specific compliant aorta replica using Shake-the-Box

High-fidelity in vitro flow simulator in combination with high-dimensional flow visualization techniques can offer precise and comprehensive evaluation of aortic hemodynamics. However, it is particularly challenging to create a fully transparent aorta replica that faithfully mimics the aortic curvature and stiffness. In this study, we successfully manufactured a patient-specific compliant aorta phantom with a dilated ascending aorta that can be used in vitro hemodynamic study. We conducted pulsatile flow measurement on the deformable aorta replica using advanced 4D particle tracking velocimetry - Shake-the-Box. The aortic distensibility, circumferential strain, flow pattern, wall shear stress (WSS), and turbulent kinetic energy were assessed. Furthermore, the peak velocity field and WSS distribution were compared to in vivo MRI measurements. We found that the distensibility and circumferential strain of our aortic replica fell within the physiological range of young patients. The aortic diameter changed as much as 5.4 mm (42 %) in a cardiac cycle and the aortic distensibility was 9.9 × 10<sup>-3</sup> mmHg<sup>-1</sup>. In addition, the obtained flow pattern and WSS distribution were found in a good agreement with in vivo MRI measurement. In conclusion, the compliant aorta phantom replicated the aortic wall material well. It also faithfully simulated the aortic flow and near-wall hemodynamics. The relatively large lumen dimension change (5.4 mm) in a cardiac cycle suggests the necessity of considering wall deformation in aortic flow simulations. We propose employing this approach for future studies, such as medical treatment training, validation of in silico fluid-structure interaction models, or as a complement to in vivo measurements.

Authors

Institutions

Publication Details

Journal
Medical Engineering & Physics
Published
2024-11-26
DOI
https://doi.org/10.1016/j.medengphy.2024.104263
Citations
6
Primary Topic
Aortic Disease and Treatment Approaches
Type
article
Field-Weighted Citation Impact
2.51

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Aortic strain, flow pattern and wall shear stress in a patient-specific compliant aorta replica using Shake-the-Box

K.M.B. Jansen, Saša Kenjereš, Jos J.M. Westenberg, Hildo J. Lamb et al.
6 citations
Medical Engineering & Physics
Aortic Disease and Treatment Approaches
2.51
article

Aortic strain, flow pattern and wall shear stress in a patient-specific compliant aorta replica using Shake-the-Box

K.M.B. Jansen, Saša Kenjereš, Jos J.M. Westenberg, Hildo J. Lamb, Xiaolin Wu
article en
6 citations

Abstract

High-fidelity in vitro flow simulator in combination with high-dimensional flow visualization techniques can offer precise and comprehensive evaluation of aortic hemodynamics. However, it is particularly challenging to create a fully transparent aorta replica that faithfully mimics the aortic curvature and stiffness. In this study, we successfully manufactured a patient-specific compliant aorta phantom with a dilated ascending aorta that can be used in vitro hemodynamic study. We conducted pulsatile flow measurement on the deformable aorta replica using advanced 4D particle tracking velocimetry - Shake-the-Box. The aortic distensibility, circumferential strain, flow pattern, wall shear stress (WSS), and turbulent kinetic energy were assessed. Furthermore, the peak velocity field and WSS distribution were compared to in vivo MRI measurements. We found that the distensibility and circumferential strain of our aortic replica fell within the physiological range of young patients. The aortic diameter changed as much as 5.4 mm (42 %) in a cardiac cycle and the aortic distensibility was 9.9 × 10<sup>-3</sup> mmHg<sup>-1</sup>. In addition, the obtained flow pattern and WSS distribution were found in a good agreement with in vivo MRI measurement. In conclusion, the compliant aorta phantom replicated the aortic wall material well. It also faithfully simulated the aortic flow and near-wall hemodynamics. The relatively large lumen dimension change (5.4 mm) in a cardiac cycle suggests the necessity of considering wall deformation in aortic flow simulations. We propose employing this approach for future studies, such as medical treatment training, validation of in silico fluid-structure interaction models, or as a complement to in vivo measurements.

Medical Engineering & PhysicsVol. 135(1)
Leiden University Medical Center (NL), J.M. Burgerscentrum (NL), Delft University of Technology (NL)
China Scholarship Council
Openalex Percentile: Top 10%
Aortic Disease and Treatment Approaches
2.51
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.