Multi-institutional experience treating patients with cardiac devices on a 1.5 Tesla magnetic resonance-linear accelerator and workflow development for thoracic treatments
CIED patients can safely undergo treatment on a 1.5 T MR-linac following RT and MRI guidelines. For targets near CIEDs, a B<sub>0</sub>-mapping procedure was considered accurate enough to determine MRgRT eligibility. Pulse oximetry is recommended for cardiac monitoring during MRI scanning due to ECG signal distortion.