Complexation Behavior and Clinical Assessment of Isomeric Calcium Ionophores of ETH 1001 in Polymeric Ion-Selective Membranes

Calcium ions are crucial in numerous physiological processes, and their precise measurement is important for many clinical diagnostics and therapeutic interventions. Traditional detection methods, such as atomic absorption spectroscopy, cannot meet clinical requirements, as they measure total calcium instead of the clinically relevant ionized form (Ca<sup>2+</sup>). Ion-selective electrodes (ISEs) provide a convenient, accurate, and specific method for Ca<sup>2+</sup> determination. Here, two isomeric calcium ionophores (<i>R</i>,<i>R</i>)-calcium ionophore I, also known as ETH 1001, and (<i>R</i>,<i>S</i>)-calcium ionophore I are synthesized and characterized for the analysis of whole blood samples with the Eaglenos blood gas analyzer (model: EG-i30) equipped with test cartridges containing screen-printed electrodes. (<i>R</i>,<i>R</i>)-Calcium ionophore I demonstrated excellent precision in whole blood samples, achieving an average bias of -2.2% compared with the available gold standard. On the other hand, the (<i>R</i>,<i>S</i>) isomer was not satisfactory, exhibiting a bias of up to -20%. Ion transfer voltammetry at thin membrane films gave information about the complex stoichiometry, complex formation constants, and ion selectivity for the two isomeric ionophores. A stoichiometry of the Ca<sup>2+</sup>-ionophore complex was confirmed to be 1:2 for both ionophores, while the (<i>R</i>,<i>R</i>) isomer gave 3.4 orders of magnitude larger complex formation constants and a modestly higher selectivity. While these data are valuable, the poor performance of membranes containing the (<i>R</i>,<i>S</i>) isomer is not directly apparent from the fundamental binding characteristics. It may be caused by interference from lipophilic blood sample components and/or surface adsorption processes, suggesting that routine selectivity characterizations of membranes containing selective ionophores are insufficient to assess their usefulness in clinical applications.

Authors

Institutions

Publication Details

Journal
ACS Sensors
Published
2024-11-26
DOI
https://doi.org/10.1021/acssensors.4c01907
Citations
6
Primary Topic
Analytical Chemistry and Sensors
Type
article
Field-Weighted Citation Impact
1.73

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Complexation Behavior and Clinical Assessment of Isomeric Calcium Ionophores of ETH 1001 in Polymeric Ion-Selective Membranes

Gabriel J. Mattos, Ju Gao, Eric Bakker, Tianbao Yang et al.
6 citations
ACS Sensors
Analytical Chemistry and Sensors
1.73
article

Complexation Behavior and Clinical Assessment of Isomeric Calcium Ionophores of ETH 1001 in Polymeric Ion-Selective Membranes

Gabriel J. Mattos, Ju Gao, Eric Bakker, Tianbao Yang, Zejian Huang, Nikolai Yu. Tiuftiakov, Yupu Zhang, Yaotian Wu, Xue Li, Junyu Zhou, Yu Qin, Zhang Ye
article en
6 citations

Abstract

Calcium ions are crucial in numerous physiological processes, and their precise measurement is important for many clinical diagnostics and therapeutic interventions. Traditional detection methods, such as atomic absorption spectroscopy, cannot meet clinical requirements, as they measure total calcium instead of the clinically relevant ionized form (Ca<sup>2+</sup>). Ion-selective electrodes (ISEs) provide a convenient, accurate, and specific method for Ca<sup>2+</sup> determination. Here, two isomeric calcium ionophores (<i>R</i>,<i>R</i>)-calcium ionophore I, also known as ETH 1001, and (<i>R</i>,<i>S</i>)-calcium ionophore I are synthesized and characterized for the analysis of whole blood samples with the Eaglenos blood gas analyzer (model: EG-i30) equipped with test cartridges containing screen-printed electrodes. (<i>R</i>,<i>R</i>)-Calcium ionophore I demonstrated excellent precision in whole blood samples, achieving an average bias of -2.2% compared with the available gold standard. On the other hand, the (<i>R</i>,<i>S</i>) isomer was not satisfactory, exhibiting a bias of up to -20%. Ion transfer voltammetry at thin membrane films gave information about the complex stoichiometry, complex formation constants, and ion selectivity for the two isomeric ionophores. A stoichiometry of the Ca<sup>2+</sup>-ionophore complex was confirmed to be 1:2 for both ionophores, while the (<i>R</i>,<i>R</i>) isomer gave 3.4 orders of magnitude larger complex formation constants and a modestly higher selectivity. While these data are valuable, the poor performance of membranes containing the (<i>R</i>,<i>S</i>) isomer is not directly apparent from the fundamental binding characteristics. It may be caused by interference from lipophilic blood sample components and/or surface adsorption processes, suggesting that routine selectivity characterizations of membranes containing selective ionophores are insufficient to assess their usefulness in clinical applications.

ACS SensorsVol. 9(12)
University of Geneva (CH)
China Scholarship Council
Openalex Percentile: Top 17%
Analytical Chemistry and Sensors
1.73
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.