Characterization of the Dynamic Flow Response in Microfluidic Devices

The purpose of this study is to characterize the dynamic response of fluid flow in microchannels, which can show significant delay times before reaching steady flow conditions. Two main sources of these delays are numerically and experimentally investigated, the hydraulic compliance which originates from the flexibility of the system components (microchannel, tubing, syringe, etc.), and the compressibility of the liquid dead volume in the setup, also known as the "bottleneck effect". A fluid-structure interaction model is presented for the compliance of rectangular PDMS microchannels that is used to form a numerically based relation for the compliance as a function of the pressure and geometry. This relation is successfully able to predict the dynamics of the flow inside PDMS microchannels in stop-flow experiments. The time delays associated with the bottleneck effect is also shown when using different syringe volumes, microchannel resistances, and liquid types. In these tests, the bottleneck effect has a much larger effect compared to the compliance of the PDMS microchannels. This is true even when using softer PDMS by increasing the monomer-to-curing agent mixing ratio. The characterization that is presented here allows for a simple analysis of microfluidic networks using the hydraulic-circuit approach.

Authors

Institutions

Publication Details

Journal
Small Methods
Published
2024-11-26
DOI
https://doi.org/10.1002/smtd.202401773
Citations
4
Primary Topic
Microfluidic and Capillary Electrophoresis Applications
Type
article
Field-Weighted Citation Impact
0.74

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Characterization of the Dynamic Flow Response in Microfluidic Devices

Mohamed Abdelgawad, Mohammed E. Elgack
4 citations
Small Methods
Microfluidic and Capillary Electrophoresis Applications
0.74
article

Characterization of the Dynamic Flow Response in Microfluidic Devices

Mohamed Abdelgawad, Mohammed E. Elgack
article en
4 citations

Abstract

The purpose of this study is to characterize the dynamic response of fluid flow in microchannels, which can show significant delay times before reaching steady flow conditions. Two main sources of these delays are numerically and experimentally investigated, the hydraulic compliance which originates from the flexibility of the system components (microchannel, tubing, syringe, etc.), and the compressibility of the liquid dead volume in the setup, also known as the "bottleneck effect". A fluid-structure interaction model is presented for the compliance of rectangular PDMS microchannels that is used to form a numerically based relation for the compliance as a function of the pressure and geometry. This relation is successfully able to predict the dynamics of the flow inside PDMS microchannels in stop-flow experiments. The time delays associated with the bottleneck effect is also shown when using different syringe volumes, microchannel resistances, and liquid types. In these tests, the bottleneck effect has a much larger effect compared to the compliance of the PDMS microchannels. This is true even when using softer PDMS by increasing the monomer-to-curing agent mixing ratio. The characterization that is presented here allows for a simple analysis of microfluidic networks using the hydraulic-circuit approach.

Small MethodsVol. 9(7)
American University of Sharjah (AE), Assiut University (EG)
American University of Sharjah
Openalex Percentile: Top 31%
Microfluidic and Capillary Electrophoresis Applications
0.74
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.