Correlation of cross-linking as measured by the trans-vinylene index and in vitro wear of polyethylene from various acetabular liner brands

The TVI is a reliable measure of the absorbed irradiation dose and does not alter over time in vivo. The products of various brands differ by manufacturing details and consequently cross-linking characteristics. Absorption and penetration of electron radiation and gamma radiation differ, potentially leading to higher degrees of cross-linking for electron radiation. There is a non-linear, inverse correlation between TVI and in vitro wear. The wear resistance of the HXLPE with low TVI was reduced and more comparable to CPE.

Authors

Institutions

Publication Details

Journal
Bone and Joint Research
Published
2024-11-26
DOI
https://doi.org/10.1302/2046-3758.1311.bjr-2024-0203.r1
Citations
1
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.57

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Correlation of cross-linking as measured by the trans-vinylene index and in vitro wear of polyethylene from various acetabular liner brands

Emanuel Gautier, Robin Pourzal, Michel Schläppi, Thomas Imwinkelried et al.
1 citations
Bone and Joint Research
Orthopaedic implants and arthroplasty
0.57
article

Correlation of cross-linking as measured by the trans-vinylene index and in vitro wear of polyethylene from various acetabular liner brands

Emanuel Gautier, Robin Pourzal, Michel Schläppi, Thomas Imwinkelried, Roman Heuberger, Markus Fürstner, Peter M. Wahl, Andrea Pascucci, Niels Icken
article en
1 citations

Abstract

The TVI is a reliable measure of the absorbed irradiation dose and does not alter over time in vivo. The products of various brands differ by manufacturing details and consequently cross-linking characteristics. Absorption and penetration of electron radiation and gamma radiation differ, potentially leading to higher degrees of cross-linking for electron radiation. There is a non-linear, inverse correlation between TVI and in vitro wear. The wear resistance of the HXLPE with low TVI was reduced and more comparable to CPE.

Bone and Joint ResearchVol. 13(11)
University of Bern (CH), Rush University Medical Center (US), University Hospital of Bern (CH), RMS Foundation (CH), Kantonsspital Winterthur (CH), Fribourg Development Agency (CH)
Dartmouth College, Zürcher Hochschule für Angewandte Wissenschaften
Openalex Percentile: Top 27%
Orthopaedic implants and arthroplasty
0.57
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Correlation of cross-linking as measured by the trans-vinylene index and in vitro wear of polyethylene from various acetabular liner brands — Emanuel Gautier, Robin Pourzal, et al. · Bone and Joint Research (2024) | TGRS Research Map | TGRS