Clinical Implementation and Commissioning of Monte Carlo Dose Calculation Algorithm for CyberKnife Treatment Planning

The purpose of this study is to commission and evaluate the Monte Carlo (MC) dose calculation algorithm for the CyberKnife S7 Robot system (Accuray, Inc.), which utilizes a 6 - MV X - band linear accelerator on a robotic arm with a Multi Leaf Collimator (MLC). We implemented a MC dose calculation algorithm for SRS/SRT treatment planning with the CyberKnife system. This algorithm simulates energy transport and dose deposition in tissue using three independent probability distributions for initial photon characteristics, based on dosimetric data from the Accuray Precision? system. The model was optimized by adjusting maximum energy and source full width at half maximum (FWHM) to minimize discrepancies with measured tissue phantom ratios. MLC plans were calculated and irradiated on various phantoms. Dose and profile measurements were taken with a PTW farmer - type ionization chamber and PTW microdiamond chamber and EBT3 Gafchromic films, followed by comparisons to TPS doses to determine dose differences. Patient - specific quality assurance (QA) was conducted using global gamma index criteria of 2%/2 mm. The optimal parameters for maximum energy and source FWHM were found to be 6.4 MeV and 1.8 mm, respectively. Results of calculations made using these parameters agreed approximately with those actually measured to within ?1% only. Monte Carlo simulation model of MLC covering of CyberKnife is clinically acceptable but tends to underestimate the delivered dose by an average of -1.3%. After 11 cm depths the dosage difference added on further. It is pragmatically advised to rely on the Monte Carlo model with the MLC primarily in heterogeneous regions (7) such as lungs

Authors

Publication Details

Journal
International Journal of Science and Research (IJSR)
Published
2024-11-05
DOI
https://doi.org/10.21275/sr241120080907
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Clinical Implementation and Commissioning of Monte Carlo Dose Calculation Algorithm for CyberKnife Treatment Planning

Pierre Jacob
International Journal of Science and Research (IJSR)
Advanced Radiotherapy Techniques
article

Clinical Implementation and Commissioning of Monte Carlo Dose Calculation Algorithm for CyberKnife Treatment Planning

Pierre Jacob
article en

Abstract

The purpose of this study is to commission and evaluate the Monte Carlo (MC) dose calculation algorithm for the CyberKnife S7 Robot system (Accuray, Inc.), which utilizes a 6 - MV X - band linear accelerator on a robotic arm with a Multi Leaf Collimator (MLC). We implemented a MC dose calculation algorithm for SRS/SRT treatment planning with the CyberKnife system. This algorithm simulates energy transport and dose deposition in tissue using three independent probability distributions for initial photon characteristics, based on dosimetric data from the Accuray Precision? system. The model was optimized by adjusting maximum energy and source full width at half maximum (FWHM) to minimize discrepancies with measured tissue phantom ratios. MLC plans were calculated and irradiated on various phantoms. Dose and profile measurements were taken with a PTW farmer - type ionization chamber and PTW microdiamond chamber and EBT3 Gafchromic films, followed by comparisons to TPS doses to determine dose differences. Patient - specific quality assurance (QA) was conducted using global gamma index criteria of 2%/2 mm. The optimal parameters for maximum energy and source FWHM were found to be 6.4 MeV and 1.8 mm, respectively. Results of calculations made using these parameters agreed approximately with those actually measured to within ?1% only. Monte Carlo simulation model of MLC covering of CyberKnife is clinically acceptable but tends to underestimate the delivered dose by an average of -1.3%. After 11 cm depths the dosage difference added on further. It is pragmatically advised to rely on the Monte Carlo model with the MLC primarily in heterogeneous regions (7) such as lungs

International Journal of Science and Research (IJSR)Vol. 13(11)
Good health and well-being
Openalex Percentile: Top 70%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.