Impact of radiation dose and iterative reconstruction (IR) level on low-contrast detectability with 4-AFC approach
This study evaluated impacts of dose and iterative reconstruction (IR) level on low-contrast detectability in images of AAPM CT Performance phantom using a 4-alternative forced choice (4-AFC) approach. Five medical physicists detected low-contrast and small objects having size of 3.0 mm with the 4-AFC method. The tests were conducted at three different radiation doses (35.8 mGy, 54.1 mGy, and 72.1 mGy) at various IR levels from 0% to 100%. The total number of 4-AFC questions was 330 questions in which each observer answered the questions in 60 minutes. Percent correct answers increase as the IR level increases from 0 to 100%. The percent correct answers also increase as the dose increases from 34.8 mGy to 72.1 mGy. 100% correct answers start at IR levels of 70, 60, and 40% for doses of 34.8, 54.1, and 72.1 mGy, respectively. Conclusions: Increasing IR level and dose improve image quality, i.e., higher average percent correct answers. At IR levels around 50%, the average percent correct is close to the maximum (around 100%) for all radiation doses. An optimal combination of IR and radiation dose can produce good image quality with lower radiation dose.
Authors
- Heri Sutanto (ORCID: https://orcid.org/0000-0001-7119-9633)
- Ariij Naufal (ORCID: https://orcid.org/0000-0002-4218-6259)
- Choirul Anam (ORCID: https://orcid.org/0000-0003-0156-6797)
- Riska Amilia (ORCID: https://orcid.org/0000-0002-7077-439X)
- Indah R. Ilham
Institutions
- Diponegoro University (ID)
Publication Details
- Journal
- International Journal of Scientific Research in Science and Technology
- Published
- 2024-11-25
- DOI
- https://doi.org/10.32628/ijsrst24116181
- Citations
- 2
- Primary Topic
- Radiation Dose and Imaging
- Type
- article
- Field-Weighted Citation Impact
- 0.81