Biosafety and efficacy evaluation of a biodegradable Zn-Cu-Mn stent: A long-term study in porcine coronary artery

In this study, biodegradable Zn-Cu-Mn alloy stents were implanted into porcine coronary artery for 18 months, and the in vivo biosafety and efficacy as well as the degradation behavior were systematically studied. Results showed a rapid endothelialization of the target vessel was achieved at 1 month post-implantation. Although the lumen diameter loss and local inflammation were observed at the early stage, the stented blood vessel could gradually recover with time. The lumen diameter was already close to normal range at 12 months, indicating good bioefficacy of the stent. No adverse effect on blood indexes or local accumulation of Zn, Cu or Mn elements were found after implantation, neither the malapposition and thrombosis were observed, which exhibited good biosafety of the stents. The stent could maintain the basic structure and mechanical integrity at 6 months, and remained only approximately 26 % of the stent volume at 18 months, suggesting a desirable degradation rate. In general, the Zn-Cu-Mn alloy stent showed great advantages and prospects in clinical application.

Authors

Institutions

Publication Details

Journal
Bioactive Materials
Published
2024-11-25
DOI
https://doi.org/10.1016/j.bioactmat.2024.11.022
Citations
6
Primary Topic
Corrosion Behavior and Inhibition
Type
article
Field-Weighted Citation Impact
0.70
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Biosafety and efficacy evaluation of a biodegradable Zn-Cu-Mn stent: A long-term study in porcine coronary artery

Pei Jia, Yan Chen, Jinzhou Zhu, Jialin Niu et al.
6 citations
Bioactive Materials
Corrosion Behavior and Inhibition
0.70
article

Biosafety and efficacy evaluation of a biodegradable Zn-Cu-Mn stent: A long-term study in porcine coronary artery

Pei Jia, Yan Chen, Jinzhou Zhu, Jialin Niu, Jimiao Jiang, Guangyin Yuan, Junfei Li, Yi Qian
article en
6 citations

Abstract

In this study, biodegradable Zn-Cu-Mn alloy stents were implanted into porcine coronary artery for 18 months, and the in vivo biosafety and efficacy as well as the degradation behavior were systematically studied. Results showed a rapid endothelialization of the target vessel was achieved at 1 month post-implantation. Although the lumen diameter loss and local inflammation were observed at the early stage, the stented blood vessel could gradually recover with time. The lumen diameter was already close to normal range at 12 months, indicating good bioefficacy of the stent. No adverse effect on blood indexes or local accumulation of Zn, Cu or Mn elements were found after implantation, neither the malapposition and thrombosis were observed, which exhibited good biosafety of the stents. The stent could maintain the basic structure and mechanical integrity at 6 months, and remained only approximately 26 % of the stent volume at 18 months, suggesting a desirable degradation rate. In general, the Zn-Cu-Mn alloy stent showed great advantages and prospects in clinical application.

Bioactive MaterialsVol. 45
Shanghai Jiao Tong University (CN), Ruijin Hospital (CN), MicroPort (China) (CN)
Zero hunger
Openalex Percentile: Top 29%
Corrosion Behavior and Inhibition
0.70
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.