Effect of phantom‐making error in the calculation of quantitative values of single‐photon emission computed tomography/computed tomography using system planar sensitivity

This study aimed to clarify the effect of the accuracy of phantom-making on quantification accuracy in the measurement of system planar sensitivity using a disc source. We measured the system planar sensitivity using the method recommended by the equipment manufacturer. A container of<sup>99m</sup>Tc solution was placed at a height of 10 cm from the surface of the collimator using styrofoam. The system planar sensitivity was the average of the values obtained from the two detectors, the measurements were taken thrice. The system planar sensitivity values obtained when the disc source size (5.0, 14.5 and 20 cm), solution volume (50, 100 and 150 ml) and radioactivity (nine points between 36 MBq and 505 MBq) changed were used to evaluate the quantification accuracy. The relative error for the 5.0 cm source significantly exceeded for the other source sizes (p < 0.05); however, no statistically significant difference was observed between the 14.5 cm and 20.0 cm sources. The smaller the solution volume, the smaller the relative error and the higher the quantification accuracy. The quantification accuracy increased as radioactivity decreased. The relative error was lowest at 36 MBq and highest at 505 MBq, with a difference of approximately 10%. Quantification accuracy improved as the size of the surface source exceeded 14.5 cm and the amount of solution decreased. Radioactivity at the time of measurement significantly affected quantification accuracy and 73-103 MBq of radioactivity was sufficient for repeatable and quantification accuracy.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2024-11-25
DOI
https://doi.org/10.1088/2057-1976/ad91bc
Citations
1
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.40
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Effect of phantom‐making error in the calculation of quantitative values of single‐photon emission computed tomography/computed tomography using system planar sensitivity

Y Nojima, Y. Yamazaki
1 citations
Biomedical Physics & Engineering Express
Medical Imaging Techniques and Applications
0.40
article

Effect of phantom‐making error in the calculation of quantitative values of single‐photon emission computed tomography/computed tomography using system planar sensitivity

Y Nojima, Y. Yamazaki
article en
1 citations

Abstract

This study aimed to clarify the effect of the accuracy of phantom-making on quantification accuracy in the measurement of system planar sensitivity using a disc source. We measured the system planar sensitivity using the method recommended by the equipment manufacturer. A container of<sup>99m</sup>Tc solution was placed at a height of 10 cm from the surface of the collimator using styrofoam. The system planar sensitivity was the average of the values obtained from the two detectors, the measurements were taken thrice. The system planar sensitivity values obtained when the disc source size (5.0, 14.5 and 20 cm), solution volume (50, 100 and 150 ml) and radioactivity (nine points between 36 MBq and 505 MBq) changed were used to evaluate the quantification accuracy. The relative error for the 5.0 cm source significantly exceeded for the other source sizes (p < 0.05); however, no statistically significant difference was observed between the 14.5 cm and 20.0 cm sources. The smaller the solution volume, the smaller the relative error and the higher the quantification accuracy. The quantification accuracy increased as radioactivity decreased. The relative error was lowest at 36 MBq and highest at 505 MBq, with a difference of approximately 10%. Quantification accuracy improved as the size of the surface source exceeded 14.5 cm and the amount of solution decreased. Radioactivity at the time of measurement significantly affected quantification accuracy and 73-103 MBq of radioactivity was sufficient for repeatable and quantification accuracy.

Biomedical Physics & Engineering ExpressVol. 11(1)
Niigata University (JP)
Openalex Percentile: Top 32%
Medical Imaging Techniques and Applications
0.40
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.