Safety and Efficacy of a Developed ‘Patent Ductus Arteriosus Occluder Device’: A Porcine Model Analysis Over Multiple Time Points

This study assessed the safety and performance of a patent ductus arteriosus (PDA) occluder using a porcine carotid artery model involving three male pigs. Each pig underwent a thorough physical examination and met the criteria for stable health and body weight. On day 0, the occluder was implanted under proper analgesia and anesthesia, with heparin used to prevent clot formation. The deployment process was meticulously evaluated through angiography and Quantitative Vascular Analysis (QVA) before and after implantation. Arteries were harvested at 90, 180, and 365 days for histopathological analysis. The occluder fulfilled all acceptance criteria, demonstrating successful deployment, trackability, and no cranial artery flow above the device. Gross necropsy showed a bulged carotid artery, suggesting a reaction to the device, while histopathological evaluations showed good endothelialization at all time points. The occluder was well-visualized during imaging, deployed easily, and could be withdrawn without issue. It effectively occluded the artery at all follow-ups without migrating from the target site. All animals survived the procedure and post- procedural period with the device intact. These results suggest that the PDA occluder is safe and effective in this model, showing promise for future human clinical applications, though further studies are needed to confirm long-term safety and efficacy.

Authors

Publication Details

Journal
International Journal of Innovative Science and Research Technology (IJISRT)
Published
2024-11-25
DOI
https://doi.org/10.38124/ijisrt/ijisrt24nov812
Primary Topic
Cardiovascular Conditions and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Safety and Efficacy of a Developed ‘Patent Ductus Arteriosus Occluder Device’: A Porcine Model Analysis Over Multiple Time Points

Minocha Pramod Kumar, Sharma Rahul, Kothwala Dr. Deveshkumar, Pandya Kamna et al.
International Journal of Innovative Science and Research Technology (IJISRT)
Cardiovascular Conditions and Treatments
article

Safety and Efficacy of a Developed ‘Patent Ductus Arteriosus Occluder Device’: A Porcine Model Analysis Over Multiple Time Points

Minocha Pramod Kumar, Sharma Rahul, Kothwala Dr. Deveshkumar, Pandya Kamna, Mistry Saurabh Rana Nirav
article en

Abstract

This study assessed the safety and performance of a patent ductus arteriosus (PDA) occluder using a porcine carotid artery model involving three male pigs. Each pig underwent a thorough physical examination and met the criteria for stable health and body weight. On day 0, the occluder was implanted under proper analgesia and anesthesia, with heparin used to prevent clot formation. The deployment process was meticulously evaluated through angiography and Quantitative Vascular Analysis (QVA) before and after implantation. Arteries were harvested at 90, 180, and 365 days for histopathological analysis. The occluder fulfilled all acceptance criteria, demonstrating successful deployment, trackability, and no cranial artery flow above the device. Gross necropsy showed a bulged carotid artery, suggesting a reaction to the device, while histopathological evaluations showed good endothelialization at all time points. The occluder was well-visualized during imaging, deployed easily, and could be withdrawn without issue. It effectively occluded the artery at all follow-ups without migrating from the target site. All animals survived the procedure and post- procedural period with the device intact. These results suggest that the PDA occluder is safe and effective in this model, showing promise for future human clinical applications, though further studies are needed to confirm long-term safety and efficacy.

International Journal of Innovative Science and Research Technology (IJISRT)
Good health and well-being
Openalex Percentile: Top 65%
Cardiovascular Conditions and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Safety and Efficacy of a Developed ‘Patent Ductus Arteriosus Occluder Device’: A Porcine Model Analysis Over Multiple Time Points — Minocha Pramod Kumar, Sharma Rahul, et al. · International Journal of Innovative Science and Research Technology (IJISRT) (2024) | TGRS Research Map | TGRS