<i>Introduction</i>. Monte Carlo simulation codes simulating medical imaging nuclear detectors (SIMIND) are notable tools used to model nuclear medicine experiments.This study aimed to confirm the usability of SIMIND as an alternative method for nuclear medicine experiments with a cardiac phantom HL, simulating human body structures, by comparing the actual experiment data.<i>Methods</i>. A cardiac phantom HL that simulates myocardial scintigraphy using<sup>123</sup>I-meta-iodobenzylguanidine was developed, and single-photon emission computed tomography/computed tomography imaging was performed using Discovery NM/CT 670 scanner. Aside from the main-energy window(159 keV ± 10%), additional windows were set on the low(137.5 keV ± 4% ) and high(180.5 keV ± 3%)-energy sides. The simulations were performed under the same conditions as the actual experiments. Regions of interest (ROIs) were set in each organ part of the experiments and simulated data, and a polar map for the myocardial part was developed. The mean, maximum (max), and minimum (min) counts within each ROI, as well as the relative errors of each segment in the polar map, were calculated to evaluate the accuracy of the simulation.<i>Results</i>. Overall, the results were favorable with relative errors of <10% except in some areas based on the data from the main-energy window and postreconstruction. On the other hand, relative errors of >10% were found in both the low and high subenergy windows. The smallest error occurred when assessing using mean values within the ROIs. The relative error was high at the cardiac base in the polar map evaluation; however, it remained <10% from the mid to apical heart sections.<i>Conclusion</i>. SIMIND is considered an alternative method for nuclear medicine experiments using a myocardial phantom HL that closely resembles human body structures. However, caution is warranted as accuracy may decrease under specific conditions.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2024-11-22
DOI
https://doi.org/10.1088/2057-1976/ad960d
Citations
1
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.40
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validation of the SIMIND simulation code using the myocardial phantom HL

Yoshiaki Yasumoto, Hiromitsu Daisaki, Mitsuru Sato
1 citations
Biomedical Physics & Engineering Express
Medical Imaging Techniques and Applications
0.40
article

Validation of the SIMIND simulation code using the myocardial phantom HL

Yoshiaki Yasumoto, Hiromitsu Daisaki, Mitsuru Sato
article en
1 citations

Abstract

<i>Introduction</i>. Monte Carlo simulation codes simulating medical imaging nuclear detectors (SIMIND) are notable tools used to model nuclear medicine experiments.This study aimed to confirm the usability of SIMIND as an alternative method for nuclear medicine experiments with a cardiac phantom HL, simulating human body structures, by comparing the actual experiment data.<i>Methods</i>. A cardiac phantom HL that simulates myocardial scintigraphy using<sup>123</sup>I-meta-iodobenzylguanidine was developed, and single-photon emission computed tomography/computed tomography imaging was performed using Discovery NM/CT 670 scanner. Aside from the main-energy window(159 keV ± 10%), additional windows were set on the low(137.5 keV ± 4% ) and high(180.5 keV ± 3%)-energy sides. The simulations were performed under the same conditions as the actual experiments. Regions of interest (ROIs) were set in each organ part of the experiments and simulated data, and a polar map for the myocardial part was developed. The mean, maximum (max), and minimum (min) counts within each ROI, as well as the relative errors of each segment in the polar map, were calculated to evaluate the accuracy of the simulation.<i>Results</i>. Overall, the results were favorable with relative errors of <10% except in some areas based on the data from the main-energy window and postreconstruction. On the other hand, relative errors of >10% were found in both the low and high subenergy windows. The smallest error occurred when assessing using mean values within the ROIs. The relative error was high at the cardiac base in the polar map evaluation; however, it remained <10% from the mid to apical heart sections.<i>Conclusion</i>. SIMIND is considered an alternative method for nuclear medicine experiments using a myocardial phantom HL that closely resembles human body structures. However, caution is warranted as accuracy may decrease under specific conditions.

Biomedical Physics & Engineering ExpressVol. 11(1)
Saitama Municipal Hospital (JP), Gunma Prefectural College of Health Sciences (JP)
Good health and well-being
Openalex Percentile: Top 32%
Medical Imaging Techniques and Applications
0.40
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.