Enhanced Effect of Patient Room Disinfection Against Carbapenem-Resistant Enterobacter cloacae and Methicillin-Resistant Staphylococcus aureus Using UV-C Irradiation in Conjunction with UV-C Containment Unit

<b>Background/Objectives</b>: In environments with high-frequency contact surfaces, drug-resistant bacteria, such as carbapenem-resistant <i>Enterobacterales</i> and methicillin-resistant <i>Staphylococcus aureus</i> (MRSA), can survive for extended periods, contributing to healthcare-associated infections. Ultraviolet (UV)-C irradiation often fails to adequately disinfect shadowed areas, leading to a persistent contamination risk. We evaluated the effectiveness of using a UV-C containment unit (UVCCU) in conjunction with UV-C irradiation to improve the sterilization effects on both direct and indirect surfaces, including shadowed areas, and to assess the leakage of UV radiation to the surroundings. <b>Methods</b>: In a model patient room, agar media inoculated with carbapenem-resistant <i>Enterobacter cloacae</i> and MRSA were placed at multiple locations on direct and indirect surfaces around the bed. We used the UV-C irradiation system, UVDI-360, to irradiate the bedroom-environment surfaces with and without a UVCCU. The reduction in bacterial colony counts with and without the UVCCU was measured by counting colony-forming units and calculating the log reduction values, and the UV radiation leakage outside the UVCCU was measured. <b>Results</b>: The use of the UVCCU led to a significant reduction in MRSA colony counts, even in shadowed areas that had previously been inadequately disinfected (with the UVCCU: 2.7 [2.7-2.8]; without the UVCCU: 0.6 [0.5-0.7]; <i>p</i> < 0.01). Additionally, the use of the UVCCU kept the UV radiation leakage to the surrounding environment within regulated limits. <b>Conclusions</b>: These findings suggest that a UVCCU can enhance the disinfection efficacy for multidrug-resistant organisms on healthcare environmental surfaces. The portability and ease of use of the UVCCU indicate its promise as an auxiliary device for UV-C disinfection in healthcare settings.

Authors

Institutions

Publication Details

Journal
Antibiotics
Published
2024-11-22
DOI
https://doi.org/10.3390/antibiotics13121115
Primary Topic
Infection Control in Healthcare
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Enhanced Effect of Patient Room Disinfection Against Carbapenem-Resistant Enterobacter cloacae and Methicillin-Resistant Staphylococcus aureus Using UV-C Irradiation in Conjunction with UV-C Containment Unit

Yoshitomo Honda, Risako Kakuta, Kentaro Takei, Shiori Kitaya et al.
Antibiotics
Infection Control in Healthcare
article

Enhanced Effect of Patient Room Disinfection Against Carbapenem-Resistant Enterobacter cloacae and Methicillin-Resistant Staphylococcus aureus Using UV-C Irradiation in Conjunction with UV-C Containment Unit

Yoshitomo Honda, Risako Kakuta, Kentaro Takei, Shiori Kitaya, Hajime Kanamori
article en

Abstract

<b>Background/Objectives</b>: In environments with high-frequency contact surfaces, drug-resistant bacteria, such as carbapenem-resistant <i>Enterobacterales</i> and methicillin-resistant <i>Staphylococcus aureus</i> (MRSA), can survive for extended periods, contributing to healthcare-associated infections. Ultraviolet (UV)-C irradiation often fails to adequately disinfect shadowed areas, leading to a persistent contamination risk. We evaluated the effectiveness of using a UV-C containment unit (UVCCU) in conjunction with UV-C irradiation to improve the sterilization effects on both direct and indirect surfaces, including shadowed areas, and to assess the leakage of UV radiation to the surroundings. <b>Methods</b>: In a model patient room, agar media inoculated with carbapenem-resistant <i>Enterobacter cloacae</i> and MRSA were placed at multiple locations on direct and indirect surfaces around the bed. We used the UV-C irradiation system, UVDI-360, to irradiate the bedroom-environment surfaces with and without a UVCCU. The reduction in bacterial colony counts with and without the UVCCU was measured by counting colony-forming units and calculating the log reduction values, and the UV radiation leakage outside the UVCCU was measured. <b>Results</b>: The use of the UVCCU led to a significant reduction in MRSA colony counts, even in shadowed areas that had previously been inadequately disinfected (with the UVCCU: 2.7 [2.7-2.8]; without the UVCCU: 0.6 [0.5-0.7]; <i>p</i> < 0.01). Additionally, the use of the UVCCU kept the UV radiation leakage to the surrounding environment within regulated limits. <b>Conclusions</b>: These findings suggest that a UVCCU can enhance the disinfection efficacy for multidrug-resistant organisms on healthcare environmental surfaces. The portability and ease of use of the UVCCU indicate its promise as an auxiliary device for UV-C disinfection in healthcare settings.

AntibioticsVol. 13(12)
Kanazawa University (JP), Tohoku University (JP)
Life in Land
Openalex Percentile: Top 74%
Infection Control in Healthcare
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.