Validation of the FF680 upper-arm blood pressure monitor according to the ISO 81060-2:2018

This study evaluated the accuracy of blood pressure measurement in adults using FF680 electronic blood pressure monitors with the Korotkoff-Sound method, in accordance with the ISO 81060-2:2018. The study was conducted at Shijiazhuang People's Hospital and lasted 56 days, from 26 May to 21 July 2023. Participants, drawn from an adult demographic, underwent data verification and analysis with strict adherence to the trial protocol. For the FF680 electronic blood pressure monitor using the Korotkoff-Sound method, data from 85 valid participants were analyzed. The findings revealed mean differences (SDs) of -0.66 mmHg (2.45 mmHg) for SBP and -0.28 mmHg (2.09 mmHg) for DBP. SBP and DBP had SDs of ≤6.9 and ≤6.95 mmHg, meeting the standard requirements. FF680 devices are recommended for adult blood pressure monitoring because they meet the ISO 81060-2:2018.

Authors

Institutions

Publication Details

Journal
Blood Pressure Monitoring
Published
2024-11-21
DOI
https://doi.org/10.1097/mbp.0000000000000735
Citations
2
Primary Topic
Blood Pressure and Hypertension Studies
Type
article
Field-Weighted Citation Impact
0.80
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validation of the FF680 upper-arm blood pressure monitor according to the ISO 81060-2:2018

Hao Chen
2 citations
Blood Pressure Monitoring
Blood Pressure and Hypertension Studies
0.80
article

Validation of the FF680 upper-arm blood pressure monitor according to the ISO 81060-2:2018

Hao Chen
article en
2 citations

Abstract

This study evaluated the accuracy of blood pressure measurement in adults using FF680 electronic blood pressure monitors with the Korotkoff-Sound method, in accordance with the ISO 81060-2:2018. The study was conducted at Shijiazhuang People's Hospital and lasted 56 days, from 26 May to 21 July 2023. Participants, drawn from an adult demographic, underwent data verification and analysis with strict adherence to the trial protocol. For the FF680 electronic blood pressure monitor using the Korotkoff-Sound method, data from 85 valid participants were analyzed. The findings revealed mean differences (SDs) of -0.66 mmHg (2.45 mmHg) for SBP and -0.28 mmHg (2.09 mmHg) for DBP. SBP and DBP had SDs of ≤6.9 and ≤6.95 mmHg, meeting the standard requirements. FF680 devices are recommended for adult blood pressure monitoring because they meet the ISO 81060-2:2018.

Blood Pressure MonitoringVol. 30(3)
First Hospital of Shijiazhuang (CN)
Partnerships for the goals
Openalex Percentile: Top 22%
Blood Pressure and Hypertension Studies
0.80
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Validation of the FF680 upper-arm blood pressure monitor according to the ISO 81060-2:2018 — Hao Chen · Blood Pressure Monitoring (2024) | TGRS Research Map | TGRS