Validation of the G.LAB MD6300 upper arm blood pressure monitor in different positions according to the AAMI/ESH/ISO Universal Standard (ISO81060-2:2018+AMD1:2020)
The G.LAB MD6300 upper arm blood pressure monitor fulfilled the validation criteria of the AAMI/ESH/ISO Universal Standard (ISO 81060-2 : 2018+AMD1 : 2020) in the sitting position, and its accuracy in the supine and right lateral decubitus position was acceptable and roughly equivalent to that in the sitting position. Therefore, it can be recommended for clinical use and self-measurement in the general population.