Innovative Approaches to Industrial Odour Monitoring: From Chemical Analysis to Predictive Models

This study evaluated the reliability of an electronic nose in monitoring odour concentration near a wastewater treatment plant and examined the correlation between four sensor readings and odour intensity. The electronic nose chemical sensors are related to the concentration of the following chemical species: two values for the concentration of VOCs recorded via the PID sensor (VPID) and the EC sensor (VEC), and concentrations of sulfuric acid (VH2S) and benzene (VC6H6). Using Random Forest and least squares regression analysis, the study identifies VH2S and VC6H6 as key contributors to odour concentration (CcOD). Three Random Forest models (RF0, RF1, RF2), with different characteristics for splitting between the test set and the training set, were tested, with RF1 showing superior predictive performance due to its training approach. All models highlighted VH2S and VC6H6 as significant predictors, while VPID and VEC had less influence. A significant correlation between odour concentration and specific chemical sensor readings was found, particularly for VH2S and VC6H6. However, predicting odour concentrations below 1000 ouE/m3 proved challenging. Linear regression further confirmed the importance of VH2S and VC6H6, with a moderate R-squared value of 0.70, explaining 70% of the variability in odour concentration. The study demonstrated the effectiveness of combining Random Forest and least squares regression for robust and interpretable results. Future research should focus on expanding the dataset and incorporating additional variables to enhance model accuracy. The findings underscore the necessity of specific sensor training and standardised procedures for accurate odour monitoring and characterisation.

Authors

Institutions

Publication Details

Journal
Atmosphere
Published
2024-11-21
DOI
https://doi.org/10.3390/atmos15121401
Citations
3
Primary Topic
Advanced Chemical Sensor Technologies
Type
article
Field-Weighted Citation Impact
0.55
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Innovative Approaches to Industrial Odour Monitoring: From Chemical Analysis to Predictive Models

Domenico Cipriano, Corrado Corradi, Ezio Bolzacchini, Dario Ronzio et al.
3 citations
Atmosphere
Advanced Chemical Sensor Technologies
0.55
article

Innovative Approaches to Industrial Odour Monitoring: From Chemical Analysis to Predictive Models

Domenico Cipriano, Corrado Corradi, Ezio Bolzacchini, Dario Ronzio, Amedeo Manuel Cefalì, Luca Ferrero, Claudia Franchina, Martina Gianotti, Alessandro Frugis, Giulio De Prosperis
article en
3 citations

Abstract

This study evaluated the reliability of an electronic nose in monitoring odour concentration near a wastewater treatment plant and examined the correlation between four sensor readings and odour intensity. The electronic nose chemical sensors are related to the concentration of the following chemical species: two values for the concentration of VOCs recorded via the PID sensor (VPID) and the EC sensor (VEC), and concentrations of sulfuric acid (VH2S) and benzene (VC6H6). Using Random Forest and least squares regression analysis, the study identifies VH2S and VC6H6 as key contributors to odour concentration (CcOD). Three Random Forest models (RF0, RF1, RF2), with different characteristics for splitting between the test set and the training set, were tested, with RF1 showing superior predictive performance due to its training approach. All models highlighted VH2S and VC6H6 as significant predictors, while VPID and VEC had less influence. A significant correlation between odour concentration and specific chemical sensor readings was found, particularly for VH2S and VC6H6. However, predicting odour concentrations below 1000 ouE/m3 proved challenging. Linear regression further confirmed the importance of VH2S and VC6H6, with a moderate R-squared value of 0.70, explaining 70% of the variability in odour concentration. The study demonstrated the effectiveness of combining Random Forest and least squares regression for robust and interpretable results. Future research should focus on expanding the dataset and incorporating additional variables to enhance model accuracy. The findings underscore the necessity of specific sensor training and standardised procedures for accurate odour monitoring and characterisation.

AtmosphereVol. 15(12)
Ricerca sul Sistema Energetico (Italy) (IT), University of Milano-Bicocca (IT)
Clean water and sanitation
Openalex Percentile: Top 36%
Advanced Chemical Sensor Technologies
0.55
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.