The impact of tube voltage on the erosion of rotating x‐ray anodes

Correcting the x-ray conversion efficiency of eroded target material, that is typically not accessible by measuring the tube current, may imply corrections to existing x-ray dose calculations. The relative increase of the allowable anode input power of rotating anode x-ray tubes with increasing tube voltage is substantially smaller than predicted by volume heating models that only rely on the focal spot surface temperature. The widely used voltage agnostic Müller-Oosterkamp formalism fails to predict the tube voltage dependency of the surface temperature of rotating anode targets, ignores the temperature dependency of the thermal diffusivity of tungsten-rhenium, and the granularity of the material. Nevertheless, we show theoretically why, backed by experience, the practical use of the Müller-Oosterkamp formalism is justified in medical imaging and provides a basis for comparison with new microparticle based targets. The reason for this surprising finding is that voltage dependent material erosion must be primarily considered as a precursor of thermal runaway effects.

Authors

Institutions

Publication Details

Journal
Medical Physics
Published
2024-11-21
DOI
https://doi.org/10.1002/mp.17528
Citations
6
Primary Topic
Hydrogen embrittlement and corrosion behaviors in metals
Type
article
Field-Weighted Citation Impact
1.03
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The impact of tube voltage on the erosion of rotating x‐ray anodes

Christopher Hulme-Smith, P. Tolias, Mats Danielsson, Rolf Behling
6 citations
Medical Physics
Hydrogen embrittlement and corrosion behaviors in metals
1.03
article

The impact of tube voltage on the erosion of rotating x‐ray anodes

Christopher Hulme-Smith, P. Tolias, Mats Danielsson, Rolf Behling
article en
6 citations

Abstract

Correcting the x-ray conversion efficiency of eroded target material, that is typically not accessible by measuring the tube current, may imply corrections to existing x-ray dose calculations. The relative increase of the allowable anode input power of rotating anode x-ray tubes with increasing tube voltage is substantially smaller than predicted by volume heating models that only rely on the focal spot surface temperature. The widely used voltage agnostic Müller-Oosterkamp formalism fails to predict the tube voltage dependency of the surface temperature of rotating anode targets, ignores the temperature dependency of the thermal diffusivity of tungsten-rhenium, and the granularity of the material. Nevertheless, we show theoretically why, backed by experience, the practical use of the Müller-Oosterkamp formalism is justified in medical imaging and provides a basis for comparison with new microparticle based targets. The reason for this surprising finding is that voltage dependent material erosion must be primarily considered as a precursor of thermal runaway effects.

Medical PhysicsVol. 52(2)
KTH Royal Institute of Technology (SE)
Life in Land
Openalex Percentile: Top 29%
Hydrogen embrittlement and corrosion behaviors in metals
1.03
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.