The hemolysis index as a tool for monitoring mild hemolysis in biochemical assays at the emergency laboratory

Hemolysis is a common pre-analytical error that can affect the accuracy of biochemical assay results. The aim of the study is to investigate the impact of hemolysis on the measurement of nine emergency biochemical parameters: glucose, creatinine, calcium, potassium, sodium, Aspartate Aminotransferase (AST), Alanine Aminotransferase (ALT),total bilirubin, and direct bilirubin. The study included blood samples collected from various units of the maternity hospital, including high-risk pregnancies follow-up, postpartum care, gynecology, neonatology, and intensive care. The study employs a Hemolysis Index (HI) to monitor and assess sample quality, determining their suitability for the analysis of the nine aforementioned biochemical parameters. The results reveal that hemolysis can significantly affect the accuracy of tests results for the studied parameters, particularly for potassium (increased 4-12 times) and AST (increased 1.47-48.55 times). Three of the studied substances showed a significant, negative, and moderate correlation with HI after the osmotic shock was induced, including AST (r = -0.501, p = 0.006), ALT (r = -0.516, p = 0.004), and sodium (r = -0.598, p = 0.001). Conversely, no significant association was found for the other parameters, which are: glucose (r = 0.079, p = 0.639); creatinine (r = 0.140, p = 0.402); direct bilirubin (r = 0.292, p = 0.075); total bilirubin (r = 0.272, p = 0.114); calcium (r = 0.215, p = 0.196); and potassium (r = -0.188, p = 0.258). Our findings indicate that HI may not be helpful for calculating a predicted value for samples with HI ranging between 87.30 and 295.9. Therefore, it is crucial to establish a threshold for the degree of hemolysis beyond which releasing the result would be considered potentially harmful to the patient especially new-borns and premature infants. In conclusion, the appropriate use of HI in clinical laboratories can enhance patient care quality by minimizing the risk of misdiagnosis leading to inappropriate treatment, particularly when requesting a second sample is not feasible.

Authors

Institutions

Publication Details

Journal
Journal of Biological Research - Bollettino della Società Italiana di Biologia Sperimentale
Published
2024-11-21
DOI
https://doi.org/10.4081/jbr.2024.12469
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The hemolysis index as a tool for monitoring mild hemolysis in biochemical assays at the emergency laboratory

Abbassia Demmouche, Sofiane Bouazza, Slimane Brikhou, Ali Khalfa et al.
Journal of Biological Research - Bollettino della Società Italiana di Biologia Sperimentale
Clinical Laboratory Practices and Quality Control
article

The hemolysis index as a tool for monitoring mild hemolysis in biochemical assays at the emergency laboratory

Abbassia Demmouche, Sofiane Bouazza, Slimane Brikhou, Ali Khalfa, Amine Bouarfa
article en

Abstract

Hemolysis is a common pre-analytical error that can affect the accuracy of biochemical assay results. The aim of the study is to investigate the impact of hemolysis on the measurement of nine emergency biochemical parameters: glucose, creatinine, calcium, potassium, sodium, Aspartate Aminotransferase (AST), Alanine Aminotransferase (ALT),total bilirubin, and direct bilirubin. The study included blood samples collected from various units of the maternity hospital, including high-risk pregnancies follow-up, postpartum care, gynecology, neonatology, and intensive care. The study employs a Hemolysis Index (HI) to monitor and assess sample quality, determining their suitability for the analysis of the nine aforementioned biochemical parameters. The results reveal that hemolysis can significantly affect the accuracy of tests results for the studied parameters, particularly for potassium (increased 4-12 times) and AST (increased 1.47-48.55 times). Three of the studied substances showed a significant, negative, and moderate correlation with HI after the osmotic shock was induced, including AST (r = -0.501, p = 0.006), ALT (r = -0.516, p = 0.004), and sodium (r = -0.598, p = 0.001). Conversely, no significant association was found for the other parameters, which are: glucose (r = 0.079, p = 0.639); creatinine (r = 0.140, p = 0.402); direct bilirubin (r = 0.292, p = 0.075); total bilirubin (r = 0.272, p = 0.114); calcium (r = 0.215, p = 0.196); and potassium (r = -0.188, p = 0.258). Our findings indicate that HI may not be helpful for calculating a predicted value for samples with HI ranging between 87.30 and 295.9. Therefore, it is crucial to establish a threshold for the degree of hemolysis beyond which releasing the result would be considered potentially harmful to the patient especially new-borns and premature infants. In conclusion, the appropriate use of HI in clinical laboratories can enhance patient care quality by minimizing the risk of misdiagnosis leading to inappropriate treatment, particularly when requesting a second sample is not feasible.

Journal of Biological Research - Bollettino della Società Italiana di Biologia Sperimentale
Université Oran 1 Ahmed Ben Bella (DZ), Université Djilali de Sidi Bel Abbès (DZ), Université de ain Témouchent (DZ)
Good health and well-being
Openalex Percentile: Top 70%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.