Feasibility study of modularized pin ridge filter implementation in proton FLASH planning for liver stereotactic ablative body radiotherapy
<i>Objective.</i>We previously developed a FLASH planning framework for streamlined pin-ridge-filter (pin-RF) design, demonstrating its feasibility for single-energy proton FLASH planning. In this study, we refined the pin-RF design for easy assembly using reusable modules, focusing on its application in liver stereotactic ablative body radiotherapy (SABR).<i>Approach.</i>This framework generates an intermediate intensity-modulated proton therapy (IMPT) plan and translates it into step widths and thicknesses of pin-RFs for a single-energy FLASH plan. Parameters like energy spacing, monitor unit limit, and spot quantity were adjusted during IMPT planning, resulting in pin-RFs assembled using predefined modules with widths from 1 to 6 mm, each with a water-equivalent-thickness of 5 mm. This approach was validated on three liver SABR cases. FLASH doses, quantified using the FLASH effectiveness model at 1-5 Gy thresholds, were compared to conventional IMPT (IMPT-CONV) doses to assess clinical benefits.<i>Main results.</i>The highest demand for 6 mm width modules, moderate for 2-4 mm, and minimal for 1- and 5-mm modules were shown across all cases. At lower dose thresholds, the two-beam case reduced indicators including liver<i>V</i><sub>21Gy</sub>and skin<i>D</i><sub>max</sub>by >19.4%, while the three-beam cases showed reductions⩽11.4%, indicating the need for higher fractional beam doses for an enhanced FLASH effect. Positive clinical benefits were seen only in the two-beam case at the 5 Gy threshold. At the 1 Gy threshold, the two-beam FLASH plan outperformed the IMPT-CONV plan, reducing dose indicators for all relevant normal tissues by up to 31.2%. In contrast, the three-beam cases showed negative clinical benefits, with skin<i>D</i><sub>max</sub>and liver<i>V</i><sub>21Gy</sub>increasing by up to 17.4% due to lower fractional beam doses and closer beam arrangements.<i>Significance.</i>This study evaluated the feasibility of modularizing streamlined pin-RFs in single-energy proton FLASH planning for liver SABR, offering guidance on optimal module composition and strategies to enhance FLASH planning.
Authors
- Jun Zhou (ORCID: https://orcid.org/0000-0002-6078-9424)
- Jufri Setianegara (ORCID: https://orcid.org/0000-0002-0919-1961)
- David S. Yu (ORCID: https://orcid.org/0000-0002-2781-5379)
- Chaoqiong Ma (ORCID: https://orcid.org/0000-0002-2276-0778)
- Yuan Gao (ORCID: https://orcid.org/0000-0002-6988-4217)
- Xiaofeng Yang (ORCID: https://orcid.org/0000-0003-4825-5515)
- Pretesh Patel
- Yinan Wang (ORCID: https://orcid.org/0009-0009-2134-712X)
Institutions
- Emory Healthcare (US)
- Emory University (US)
- Emory University Hospital (US)
- Piedmont Cancer Institute (US)
- University of Kansas Medical Center (US)
Publication Details
- Journal
- Physics in Medicine and Biology
- Published
- 2024-11-21
- DOI
- https://doi.org/10.1088/1361-6560/ad95d6
- Citations
- 2
- Primary Topic
- Radiation Therapy and Dosimetry
- Type
- article
- Field-Weighted Citation Impact
- 0.84
Funders
- National Institutes of Health