Interlaboratory comparisons of the physical properties of X-ray systems and display devices in clinical conditions: whether it relates to how well we measure

Abstract Introduction: Polish Atomic Law requires that organizations conducting quality control checks on medical radiological equipment must participate in interlaboratory comparisons. Medical physicists who conduct these tests in their hospitals seek to validate their results by participating in interlaboratory comparisons (ILC) at least once every four years. Material and methods: Comparisons involved measurements of physical parameters for various medical imaging equipment, including a digital X-ray unit, an angiograph, a computed tomography scanner, and a display device used to present medical images. The reference value was defined as the weighted average of the measurements performed by the participants. The En index calculation was used to assess the measurement results of each participant individually. Results: All laboratories achieved 100% satisfactory results for computed tomography, angiography, and X-ray diagnostics, despite using mostly different research methods and various measuring equipment. During the ILC, we observed discrepancies in the results of Hounsfield Unit (HU) measurements in computed tomography (CT), which were due to the size and position of the region of interest (ROI). Additionally, there were discrepancies in the results of dose measurements in angiography, attributed to factors such as the influence of the patient’s table on the measurement and the value of the backscatter factor. As a result, a standard measurement method was established. For display device measurements, two laboratories achieved 100%, two achieved over 94%, and one achieved over 88%. Conclusions: The goals for conducting comparative research in the field of X-ray diagnostics have been met. The participants’ proper use of the measuring equipment has been confirmed, and the suitability of the research methods for the intended purpose has been verified. We identify potential sources of discrepancies during ILC conducted in clinical settings. Selecting the appropriate measurement geometry and testing procedures is crucial for obtaining valuable ILC results.

Authors

Institutions

Publication Details

Journal
Polish Journal of Medical Physics And Engineering
Published
2024-11-07
DOI
https://doi.org/10.2478/pjmpe-2024-0024
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Interlaboratory comparisons of the physical properties of X-ray systems and display devices in clinical conditions: whether it relates to how well we measure

Joanna Kidoń-Szołtysek, Paweł Wołowiec, Joanna Gaweł
Polish Journal of Medical Physics And Engineering
Radiation Dose and Imaging
article

Interlaboratory comparisons of the physical properties of X-ray systems and display devices in clinical conditions: whether it relates to how well we measure

Joanna Kidoń-Szołtysek, Paweł Wołowiec, Joanna Gaweł
article en

Abstract

Abstract Introduction: Polish Atomic Law requires that organizations conducting quality control checks on medical radiological equipment must participate in interlaboratory comparisons. Medical physicists who conduct these tests in their hospitals seek to validate their results by participating in interlaboratory comparisons (ILC) at least once every four years. Material and methods: Comparisons involved measurements of physical parameters for various medical imaging equipment, including a digital X-ray unit, an angiograph, a computed tomography scanner, and a display device used to present medical images. The reference value was defined as the weighted average of the measurements performed by the participants. The En index calculation was used to assess the measurement results of each participant individually. Results: All laboratories achieved 100% satisfactory results for computed tomography, angiography, and X-ray diagnostics, despite using mostly different research methods and various measuring equipment. During the ILC, we observed discrepancies in the results of Hounsfield Unit (HU) measurements in computed tomography (CT), which were due to the size and position of the region of interest (ROI). Additionally, there were discrepancies in the results of dose measurements in angiography, attributed to factors such as the influence of the patient’s table on the measurement and the value of the backscatter factor. As a result, a standard measurement method was established. For display device measurements, two laboratories achieved 100%, two achieved over 94%, and one achieved over 88%. Conclusions: The goals for conducting comparative research in the field of X-ray diagnostics have been met. The participants’ proper use of the measuring equipment has been confirmed, and the suitability of the research methods for the intended purpose has been verified. We identify potential sources of discrepancies during ILC conducted in clinical settings. Selecting the appropriate measurement geometry and testing procedures is crucial for obtaining valuable ILC results.

Polish Journal of Medical Physics And EngineeringVol. 30(4)
Medical University of Silesia (PL), Holy Cross University (PL), Silesian Center for Heart Disease (PL), University of Warsaw (PL)
Peace, Justice and strong institutions
Openalex Percentile: Top 65%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.