The enhanced sensitivity of pitot tubes at low Reynolds number

Pitot tubes are commonly used for measuring flow velocity in air and other fluids. The aim of this experimental study was to link flow conditions and geometry to the differential pressure response ( Δp ) of pitot tubes. Based on outer diameter the pitot tube Reynolds number (Re D ) ranges from less than 1 (molecular regime) to more than 100 (inertial regime). This large range in Reynolds number was achieved by varying the pitot tube outer diameter in the range 0.4–10 mm, the air density from 0.006 to 1.2 kg/m 3 and the air velocity from 0.25 to 30 m/s and by operating pitot tubes within a recirculating low pressure wind tunnel down to a pressure of 0.5 mbar as well as performing comparative observation in an open circuit wind tunnel. For the smallest Re D , viscous forces enhance Δp giving the smallest micro-pitot tubes superior functionality especially at low pressure. The small diameter of the micro-pitot tubes can also allow measurements close to a wall within the boundary layer. Measurements of Δp for Re D between approximately 0.08 and 1000 agree well with an analytical model derived by Muriel Barker in 1922 while they agree less well or even poorly with more recent models. At the lowest Re D the maximum enhancement of Δp was observed to be a factor of 65. The enhancement became undetectable for Re D larger than approximately 70. At the lowest air density and the smallest pitot tube inner diameter possible effects due to slip were observed. • Pitot tube measurements have been made at Reynolds numbers from below 0.1 to 1000. • Three different regimes of response have been identified: inertial, viscous and transitional. • Due to viscous forces the pressure coefficient c p > 1 when the pitot tube Reynolds number <70. • The output is enhanced 73 times over the Bernoulli value at the smallest pitot Reynolds number. • The observed results agree well with the theoretical predictions made by Barker (1922).

Authors

Institutions

Publication Details

Journal
Flow Measurement and Instrumentation
Published
2024-11-20
DOI
https://doi.org/10.1016/j.flowmeasinst.2024.102750
Citations
4
Primary Topic
Aerodynamics and Acoustics in Jet Flows
Type
article
Field-Weighted Citation Impact
5.20

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The enhanced sensitivity of pitot tubes at low Reynolds number

J. Iversen, K. R. Rasmussen, J. P. Merrison
4 citations
Flow Measurement and Instrumentation
Aerodynamics and Acoustics in Jet Flows
5.20
article

The enhanced sensitivity of pitot tubes at low Reynolds number

J. Iversen, K. R. Rasmussen, J. P. Merrison
article en
4 citations

Abstract

Pitot tubes are commonly used for measuring flow velocity in air and other fluids. The aim of this experimental study was to link flow conditions and geometry to the differential pressure response ( Δp ) of pitot tubes. Based on outer diameter the pitot tube Reynolds number (Re D ) ranges from less than 1 (molecular regime) to more than 100 (inertial regime). This large range in Reynolds number was achieved by varying the pitot tube outer diameter in the range 0.4–10 mm, the air density from 0.006 to 1.2 kg/m 3 and the air velocity from 0.25 to 30 m/s and by operating pitot tubes within a recirculating low pressure wind tunnel down to a pressure of 0.5 mbar as well as performing comparative observation in an open circuit wind tunnel. For the smallest Re D , viscous forces enhance Δp giving the smallest micro-pitot tubes superior functionality especially at low pressure. The small diameter of the micro-pitot tubes can also allow measurements close to a wall within the boundary layer. Measurements of Δp for Re D between approximately 0.08 and 1000 agree well with an analytical model derived by Muriel Barker in 1922 while they agree less well or even poorly with more recent models. At the lowest Re D the maximum enhancement of Δp was observed to be a factor of 65. The enhancement became undetectable for Re D larger than approximately 70. At the lowest air density and the smallest pitot tube inner diameter possible effects due to slip were observed. • Pitot tube measurements have been made at Reynolds numbers from below 0.1 to 1000. • Three different regimes of response have been identified: inertial, viscous and transitional. • Due to viscous forces the pressure coefficient c p > 1 when the pitot tube Reynolds number <70. • The output is enhanced 73 times over the Bernoulli value at the smallest pitot Reynolds number. • The observed results agree well with the theoretical predictions made by Barker (1922).

Flow Measurement and InstrumentationVol. 101
Aarhus University (DK)
European Space Agency, European Commission
Openalex Percentile: Top 5%
Aerodynamics and Acoustics in Jet Flows
5.20
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.