High-accuracy mirror surface measurement using three scanning interferometric displacement sensors

High-accuracy mirror surface measurement using three scanning interferometric displacement sensors, with motion error calibrated, is proposed for X-ray mirror metrology. The motion error of the scanning displacement stage significantly affects the high-accuracy measurement of mirror surfaces during scanning. The influence of motion errors on surface height measurements is studied. After eliminating motion errors with a system comprising three scanning interferometric displacement sensors, the mirror surface can be obtained. The formula for the surface height solution is derived, and the residual error terms are analyzed. A method of error reduction using three reference mirrors to calibrate distance is proposed. The calibration and test experiments are conducted. Experimental results demonstrate the effectiveness of calibration, and the measurement repeatability is 7.39 nm.

Authors

Institutions

Publication Details

Published
2024-11-20
DOI
https://doi.org/10.1117/12.3045125
Primary Topic
Advanced Measurement and Metrology Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

High-accuracy mirror surface measurement using three scanning interferometric displacement sensors

Guang‐Biao Zhou, Xiaohao Dong, Weizheng Lei, Jie Wang et al.
Advanced Measurement and Metrology Techniques
article

High-accuracy mirror surface measurement using three scanning interferometric displacement sensors

Guang‐Biao Zhou, Xiaohao Dong, Weizheng Lei, Jie Wang, Jiezhuo Wang
article en

Abstract

High-accuracy mirror surface measurement using three scanning interferometric displacement sensors, with motion error calibrated, is proposed for X-ray mirror metrology. The motion error of the scanning displacement stage significantly affects the high-accuracy measurement of mirror surfaces during scanning. The influence of motion errors on surface height measurements is studied. After eliminating motion errors with a system comprising three scanning interferometric displacement sensors, the mirror surface can be obtained. The formula for the surface height solution is derived, and the residual error terms are analyzed. A method of error reduction using three reference mirrors to calibrate distance is proposed. The calibration and test experiments are conducted. Experimental results demonstrate the effectiveness of calibration, and the measurement repeatability is 7.39 nm.

ShanghaiTech University (CN), Shanghai Advanced Research Institute (CN), Shanghai Institute of Applied Physics (CN), University of Chinese Academy of Sciences (CN)
Affordable and clean energy
Openalex Percentile: Top 79%
Advanced Measurement and Metrology Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

High-accuracy mirror surface measurement using three scanning interferometric displacement sensors — Guang‐Biao Zhou, Xiaohao Dong, et al. · (2024) | TGRS Research Map | TGRS