Application of improved matrix dilution method in quantitative analysis of Ni-Co-Mn ternary precursor

Abstract When using energy-dispersive X-ray fluorescence (EDXRF) to analyze Ni-Co-Mn (NCM) samples, if the standard sample's concentration greatly differs from the unknown sample's concentration, the traditional matrix dilution method requires repeated dilution and measurement. This makes the process time-consuming and labor-intensive. This study proposes an improved matrix dilution method to reduce sample preparation and analysis. This method first establishes a functional relationship model between the dilution factor and the characteristic X-ray intensity. Then the characteristic X-ray intensity of the analyzed element can be calculated by this model, avoiding unnecessary dilution and measurement steps. To verify the effectiveness of this method, the dilution factors and characteristic X-ray intensities of the test samples were fitted using the established functional relationship. The fitting results showed that the fitting coefficients of determination of the Mn, Co, and Ni were all 0.999. Quantitative analysis was performed on the characteristic X-ray intensity fitting values and measured values of the test samples. The results showed that the quantitative results of the two were consistent. The average error of the three elements for both methods was 1.1% and 0.7%, respectively. It shows that through the established functional relationship, the characteristic X-ray intensity can be effectively calculated by the dilution factor. This method can be applied to samples with identical elements and proportions of target elements, but with different concentrations, using the same set of standard samples.

Authors

Institutions

Publication Details

Journal
Nukleonika
Published
2024-11-20
DOI
https://doi.org/10.2478/nuka-2024-0027
Primary Topic
Electron and X-Ray Spectroscopy Techniques
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Application of improved matrix dilution method in quantitative analysis of Ni-Co-Mn ternary precursor

Wentao Sun, Xin Huang, Hui Li, Jianbo Yang et al.
Nukleonika
Electron and X-Ray Spectroscopy Techniques
article

Application of improved matrix dilution method in quantitative analysis of Ni-Co-Mn ternary precursor

Wentao Sun, Xin Huang, Hui Li, Jianbo Yang, Rui Li, Jie Xu, Jizhou Ren
article en

Abstract

Abstract When using energy-dispersive X-ray fluorescence (EDXRF) to analyze Ni-Co-Mn (NCM) samples, if the standard sample's concentration greatly differs from the unknown sample's concentration, the traditional matrix dilution method requires repeated dilution and measurement. This makes the process time-consuming and labor-intensive. This study proposes an improved matrix dilution method to reduce sample preparation and analysis. This method first establishes a functional relationship model between the dilution factor and the characteristic X-ray intensity. Then the characteristic X-ray intensity of the analyzed element can be calculated by this model, avoiding unnecessary dilution and measurement steps. To verify the effectiveness of this method, the dilution factors and characteristic X-ray intensities of the test samples were fitted using the established functional relationship. The fitting results showed that the fitting coefficients of determination of the Mn, Co, and Ni were all 0.999. Quantitative analysis was performed on the characteristic X-ray intensity fitting values and measured values of the test samples. The results showed that the quantitative results of the two were consistent. The average error of the three elements for both methods was 1.1% and 0.7%, respectively. It shows that through the established functional relationship, the characteristic X-ray intensity can be effectively calculated by the dilution factor. This method can be applied to samples with identical elements and proportions of target elements, but with different concentrations, using the same set of standard samples.

NukleonikaVol. 69(4)
Chengdu University of Technology (CN), Sichuan University of Science and Engineering (CN)
National Natural Science Foundation of China, Sichuan University of Science and Engineering, Sichuan University
Decent work and economic growth
Openalex Percentile: Top 80%
Electron and X-Ray Spectroscopy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.