Comparing Different Self-Tonometry Devices against Goldmann Applanation Tonometry – Special Focus on Biomechanical Properties of the Cornea

We consider differences in readings using the two self-tonometers compared against GAT as the established reference method to be within acceptable tolerances. Monitoring changes in readings using either device over an extended period - such as gradual increases in intraocular pressure values - is far more important in clinical practice. The measurement sequence requires special consideration due to interplay between measurement methods when comparing readings from different tonometers. We recommend taking iCare HOME readings before applanation tonometry. Biomechanical corneal properties especially need to be considered in iCare HOME and Ocuton S*TT-MV self-tonometers. CRF values affect readings from both self-tonometers. CH values should also be considered when interpreting readings from Ocuton S*TT-MV.

Authors

Institutions

Publication Details

Journal
Klinische Monatsblätter für Augenheilkunde
Published
2024-11-20
DOI
https://doi.org/10.1055/a-2480-1772
Primary Topic
Glaucoma and retinal disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparing Different Self-Tonometry Devices against Goldmann Applanation Tonometry – Special Focus on Biomechanical Properties of the Cornea

Frank Tost, Andreas Stahl, Allam Tayar, Rico Großjohann
Klinische Monatsblätter für Augenheilkunde
Glaucoma and retinal disorders
article

Comparing Different Self-Tonometry Devices against Goldmann Applanation Tonometry – Special Focus on Biomechanical Properties of the Cornea

Frank Tost, Andreas Stahl, Allam Tayar, Rico Großjohann
article en

Abstract

We consider differences in readings using the two self-tonometers compared against GAT as the established reference method to be within acceptable tolerances. Monitoring changes in readings using either device over an extended period - such as gradual increases in intraocular pressure values - is far more important in clinical practice. The measurement sequence requires special consideration due to interplay between measurement methods when comparing readings from different tonometers. We recommend taking iCare HOME readings before applanation tonometry. Biomechanical corneal properties especially need to be considered in iCare HOME and Ocuton S*TT-MV self-tonometers. CRF values affect readings from both self-tonometers. CH values should also be considered when interpreting readings from Ocuton S*TT-MV.

Klinische Monatsblätter für AugenheilkundeVol. 242(02)
Universitätsmedizin Greifswald (DE), Universität Greifswald (DE), Klinik und Poliklinik für Augenheilkunde Universitätsklinikum Regensburg (DE)
Openalex Percentile: Top 66%
Glaucoma and retinal disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.