Experimental test of neutron intensity monitor with isomer production reaction for p-Li neutron source for boron neutron capture therapy
Introduction Boron Neutron Capture Therapy (BNCT) is a promising cancer therapy. At present, development of accelerator based neutron source (ABNS) is underway to be utilized as a neutron source instead of nuclear reactor. However, it is known that the neutron field formed with accelerators have different characteristics depending on kinds of accelerators. We thus have to characterize the field before practical use. Method In the authors’ group, various neutronics characterization devices have been developed for our p-Li based BNCT machine named CSePT. In this paper, three neutron intensity monitor foils with an isomer production reaction for several tens to 800 keV of the p-Li neutrons were proposed, i.e., 107 Ag, 115 In and 189 Os. Result and Discussion From the experimental test results, two activation foils of 107 Ag and 115 In were confirmed to be a possible candidate as the monitor. However, the isomer production cross sections of them should be examined for practical use.
Authors
- Isao Murata (ORCID: https://orcid.org/0000-0003-2759-5561)
- Sachie Kusaka (ORCID: https://orcid.org/0009-0008-8141-7778)
- Fuminobu Sato
- Shingo Tamaki (ORCID: https://orcid.org/0000-0003-2159-1316)
- Yuichi Ohtani
Institutions
- The University of Osaka (JP)
Publication Details
- Journal
- Frontiers in Nuclear Engineering
- Published
- 2024-11-20
- DOI
- https://doi.org/10.3389/fnuen.2024.1457737
- Citations
- 1
- Primary Topic
- Nuclear Physics and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.61