Interference fit of cementless tibial implants and the initial mechanical environment: A micro‐CT and DVC study

During cementless total knee arthroplasty (TKA), an overlap between the resected tibia and the implant's geometry, termed interference fit, is introduced to facilitate primary stability and direct bone-implant contact. However, little is known about the actual interference achieved and the resulting mechanical response in the surrounding cancellous bone. The aim of this study was (1) to experimentally quantify the actual interference achieved for a commercially available cementless tibial implant and (2) to assess its effect on the post-impaction cancellous bone strain. Seven human cadaveric tibiae were micro-CT scanned intact (23 µm/pixel), once prepared for TKA (46 µm/pixel) and following implantation (46 µm/pixel). The actual interference across the entire bone-implant interface was quantified and, via digital volume correlation, the compressive strains of bone in contact with the implant and at increasing distance, were extracted. An inhomogeneous actual interference was found across the implant pegs and keel (median ± std dev: 0.70 ± 0.27 mm), which was lower than the intended. Limited interference (0.02 ± 0.12 mm) was found directly under the baseplate, with immediate bone-baseplate contact of 54%. The induced compressive strains were related to the actual interference within 3.14 mm from the bone-implant interface (R<sup>2</sup> = 0.269-0.450, p < 0.001), with higher compressive strains corresponding to higher interference, but not being related to the bone volume fraction. Clinical Significance: Insight is provided into the interaction between the variability of the resection and the resulting mechanical environment. A complex relationship is apparent, whereby the actual interference accounted for up to 45% of the variation in induced compressive strain magnitude.

Authors

Institutions

Publication Details

Journal
Journal of Orthopaedic Research®
Published
2024-11-20
DOI
https://doi.org/10.1002/jor.26003
Citations
1
Primary Topic
Total Knee Arthroplasty Outcomes
Type
article
Field-Weighted Citation Impact
0.57

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Interference fit of cementless tibial implants and the initial mechanical environment: A micro‐CT and DVC study

Mark Taylor, Egon Perilli, Sophie Rapagna, Lauren S. Wearne et al.
1 citations
Journal of Orthopaedic Research®
Total Knee Arthroplasty Outcomes
0.57
article

Interference fit of cementless tibial implants and the initial mechanical environment: A micro‐CT and DVC study

Mark Taylor, Egon Perilli, Sophie Rapagna, Lauren S. Wearne, Greg Keene
article en
1 citations

Abstract

During cementless total knee arthroplasty (TKA), an overlap between the resected tibia and the implant's geometry, termed interference fit, is introduced to facilitate primary stability and direct bone-implant contact. However, little is known about the actual interference achieved and the resulting mechanical response in the surrounding cancellous bone. The aim of this study was (1) to experimentally quantify the actual interference achieved for a commercially available cementless tibial implant and (2) to assess its effect on the post-impaction cancellous bone strain. Seven human cadaveric tibiae were micro-CT scanned intact (23 µm/pixel), once prepared for TKA (46 µm/pixel) and following implantation (46 µm/pixel). The actual interference across the entire bone-implant interface was quantified and, via digital volume correlation, the compressive strains of bone in contact with the implant and at increasing distance, were extracted. An inhomogeneous actual interference was found across the implant pegs and keel (median ± std dev: 0.70 ± 0.27 mm), which was lower than the intended. Limited interference (0.02 ± 0.12 mm) was found directly under the baseplate, with immediate bone-baseplate contact of 54%. The induced compressive strains were related to the actual interference within 3.14 mm from the bone-implant interface (R<sup>2</sup> = 0.269-0.450, p < 0.001), with higher compressive strains corresponding to higher interference, but not being related to the bone volume fraction. Clinical Significance: Insight is provided into the interaction between the variability of the resection and the resulting mechanical environment. A complex relationship is apparent, whereby the actual interference accounted for up to 45% of the variation in induced compressive strain magnitude.

Journal of Orthopaedic Research®Vol. 43(2)
Flinders University (AU), Sports Medicine Australia (AU)
Australian National Fabrication Facility, Flinders University, Australian Research Council
Openalex Percentile: Top 27%
Total Knee Arthroplasty Outcomes
0.57
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.