Gray values and noise behavior of cone-beam computed tomography machines—an <i>in vitro</i> study
The MGV and noise provided by the tested phantom vary largely among machines. The MGV is mainly influenced by the FOV size, especially for bone equivalent radiodensity. For most machines, when the acquisition parameters affect the MGV, the MGV decrease with the increase in the acquisition parameters.