Validity and Test-Retest Reliability of a Hand-Held Ultrasound Device in Measuring Kidney Metrics in Healthy Young Adults

Objective: The use of a hand-held ultrasound device provides a portable, cost-effective solution for renal assessments in rural settings, but their accuracy and reliability in measuring kidney metrics remain uncertain. The aim of this study was to evaluate the validity and reliability of the Lumify hand-held ultrasound device (LHUS) in capturing basic kidney metrics in healthy adults. Materials and Methods: Twenty individuals underwent renal sonography measurements, which included length, height, width, cortical thickness, parenchymal echogenicity, and renal resistive index (RRI). These measurements were completed using LHUS and the Epiq Elite ultrasound equipment system. LHUS measurements were repeated for test-retest reliability (TRR). Validity and TRR were assessed using intraclass correlation coefficient (ICC), Bland-Altman plots, coefficient of variation (COV), and mean absolute percent error (MAPE). Results: There was poor agreement (0.414, p = .048) between equipment systems for RRI. All other measures demonstrated moderate to excellent agreement (ICC = 0.75–0.94, p < .001) between systems. The left kidney showed poor agreement for RRI (0.06, p = .45). MAPE for all measurements between devices was <5%, indicating acceptable accuracy. All repeated measurements showed moderate to excellent reliability (ICC = 0.71–0.96, p < .001), with low mean COV for each measurement (0.02–0.08). Conclusion: Based on this study, LHUS appears to be a valid and reliable tool for basic kidney metrics in healthy adults, but caution should be used when obtaining an RRI, especially in the left kidney.

Authors

Institutions

Publication Details

Journal
Journal of diagnostic medical sonography
Published
2024-11-20
DOI
https://doi.org/10.1177/87564793241297708
Primary Topic
Renal and Vascular Pathologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validity and Test-Retest Reliability of a Hand-Held Ultrasound Device in Measuring Kidney Metrics in Healthy Young Adults

Danielle R. Bruns, Josee Walters, Evan C. Johnson, Timothy J. Robinson et al.
Journal of diagnostic medical sonography
Renal and Vascular Pathologies
article

Validity and Test-Retest Reliability of a Hand-Held Ultrasound Device in Measuring Kidney Metrics in Healthy Young Adults

Danielle R. Bruns, Josee Walters, Evan C. Johnson, Timothy J. Robinson, Emily E. Schmitt, Zachary J. Schlader, Bailee R. Smith, Jimmy Bautista, Kevin E. Miller
article en

Abstract

Objective: The use of a hand-held ultrasound device provides a portable, cost-effective solution for renal assessments in rural settings, but their accuracy and reliability in measuring kidney metrics remain uncertain. The aim of this study was to evaluate the validity and reliability of the Lumify hand-held ultrasound device (LHUS) in capturing basic kidney metrics in healthy adults. Materials and Methods: Twenty individuals underwent renal sonography measurements, which included length, height, width, cortical thickness, parenchymal echogenicity, and renal resistive index (RRI). These measurements were completed using LHUS and the Epiq Elite ultrasound equipment system. LHUS measurements were repeated for test-retest reliability (TRR). Validity and TRR were assessed using intraclass correlation coefficient (ICC), Bland-Altman plots, coefficient of variation (COV), and mean absolute percent error (MAPE). Results: There was poor agreement (0.414, p = .048) between equipment systems for RRI. All other measures demonstrated moderate to excellent agreement (ICC = 0.75–0.94, p < .001) between systems. The left kidney showed poor agreement for RRI (0.06, p = .45). MAPE for all measurements between devices was <5%, indicating acceptable accuracy. All repeated measurements showed moderate to excellent reliability (ICC = 0.71–0.96, p < .001), with low mean COV for each measurement (0.02–0.08). Conclusion: Based on this study, LHUS appears to be a valid and reliable tool for basic kidney metrics in healthy adults, but caution should be used when obtaining an RRI, especially in the left kidney.

Journal of diagnostic medical sonographyVol. 41(3)
University of Wyoming (US), University of Providence (US), Providence College (US), Indiana University Bloomington (US), Memorial Hospital (US)
No poverty
Openalex Percentile: Top 61%
Renal and Vascular Pathologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.