Around-Body Versus On-Body Motion Sensing: A Comparison of Efficacy Across a Range of Body Movements and Scales

Motion is vital for life. Currently, the clinical assessment of motion abnormalities is largely qualitative. We previously developed methods to quantitatively assess motion using visual detection systems (around-body) and stretchable electronic sensors (on-body). Here we compare the efficacy of these methods across predefined motions, hypothesizing that the around-body system detects motion with similar accuracy as on-body sensors. Six human volunteers performed six defined motions covering three excursion lengths, small, medium, and large, which were analyzed via both around-body visual marker detection (MoCa version 1.0) and on-body stretchable electronic sensors (BioStamp version 1.0). Data from each system was compared as to the extent of trackability and comparative efficacy between systems. Both systems successfully detected motions, allowing quantitative analysis. Angular displacement between systems had the highest agreement efficiency for the bicep curl and body lean motion, with 73.24% and 65.35%, respectively. The finger pinch motion had an agreement efficiency of 36.71% and chest abduction/adduction had 45.55%. Shoulder abduction/adduction and shoulder flexion/extension motions had the lowest agreement efficiencies with 24.49% and 26.28%, respectively. MoCa was comparable to BioStamp in terms of angular displacement, though velocity and linear speed output could benefit from additional processing. Our findings demonstrate comparable efficacy for non-contact motion detection to that of on-body sensor detection, and offers insight as to the best system selection for specific clinical uses based on the use-case of the desired motion being analyzed.

Authors

Institutions

Publication Details

Journal
Bioengineering
Published
2024-11-19
DOI
https://doi.org/10.3390/bioengineering11111163
Citations
1
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.18
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Around-Body Versus On-Body Motion Sensing: A Comparison of Efficacy Across a Range of Body Movements and Scales

Sara Sheikhlary, Ethan C. Ross, Rebecca C. Slepian, Zachary Hansen et al.
1 citations
Bioengineering
Non-Invasive Vital Sign Monitoring
0.18
article

Around-Body Versus On-Body Motion Sensing: A Comparison of Efficacy Across a Range of Body Movements and Scales

Sara Sheikhlary, Ethan C. Ross, Rebecca C. Slepian, Zachary Hansen, Binh Thang Tran, Marvin J. Slepian, Jordan Rodriguez, Ryan Chung, Katelyn Rohrer, Farah Alqaraghuli, Chris Hedgecoke, Youssif Abdelkeder, Camila Grubb, Luis De Anda, Mehrail Lawendy, Suleyman A. Omer, G. Pierre
article en
1 citations

Abstract

Motion is vital for life. Currently, the clinical assessment of motion abnormalities is largely qualitative. We previously developed methods to quantitatively assess motion using visual detection systems (around-body) and stretchable electronic sensors (on-body). Here we compare the efficacy of these methods across predefined motions, hypothesizing that the around-body system detects motion with similar accuracy as on-body sensors. Six human volunteers performed six defined motions covering three excursion lengths, small, medium, and large, which were analyzed via both around-body visual marker detection (MoCa version 1.0) and on-body stretchable electronic sensors (BioStamp version 1.0). Data from each system was compared as to the extent of trackability and comparative efficacy between systems. Both systems successfully detected motions, allowing quantitative analysis. Angular displacement between systems had the highest agreement efficiency for the bicep curl and body lean motion, with 73.24% and 65.35%, respectively. The finger pinch motion had an agreement efficiency of 36.71% and chest abduction/adduction had 45.55%. Shoulder abduction/adduction and shoulder flexion/extension motions had the lowest agreement efficiencies with 24.49% and 26.28%, respectively. MoCa was comparable to BioStamp in terms of angular displacement, though velocity and linear speed output could benefit from additional processing. Our findings demonstrate comparable efficacy for non-contact motion detection to that of on-body sensor detection, and offers insight as to the best system selection for specific clinical uses based on the use-case of the desired motion being analyzed.

BioengineeringVol. 11(11)
University of Arizona (US)
Openalex Percentile: Top 51%
Non-Invasive Vital Sign Monitoring
0.18
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.