An Amplitude-Programmable Energy-Recycling High-Voltage Resonant Pulser for Battery- Powered Ultrasound Devices

This article presents an application-specific integrated circuit (ASIC) for battery-powered ultrasound (US) devices. The ASIC implements a novel energy-efficient high-voltage (HV) pulser that generates HV transmit (TX) pulses directly from a low-voltage (LV) battery supply. By means of a single off-chip inductor, energy is supplied to a US transducer in a resonant fashion, directly generating half-period sinusoidal HV pulses on the transducer, while consuming substantially less energy than a conventional class-D pulser. By recycling residual reactive energy from the transducer back to the input, the energy consumption is further reduced by more than 50%. The autocalibration techniques are leveraged to deal with tolerances of the inductor, transducer, and battery supply and thus maximize the energy efficiency. A prototype chip was fabricated in TSMC 0.18-<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\mu $ </tex-math></inline-formula>m HV BCD technology and used to drive external 120-pF capacitive micromachined US transducers (CMUTs) with a center frequency of approximately 2.5 MHz. Electrical measurements show that the prototype can generate pulses with a peak amplitude between 10 and 30 V accurate to within ±1 V. Acoustic measurements demonstrate successful ultrasonic pulse transmission and pulse-echo measurements. The prototype reaches a peak efficiency of 0.23 fCV2, which is the highest reported to date for HV pulsers targeting US imaging.

Authors

Institutions

Publication Details

Journal
IEEE Journal of Solid-State Circuits
Published
2024-11-19
DOI
https://doi.org/10.1109/jssc.2024.3494536
Citations
2
Primary Topic
Ultrasound and Hyperthermia Applications
Type
article
Field-Weighted Citation Impact
0.37
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

An Amplitude-Programmable Energy-Recycling High-Voltage Resonant Pulser for Battery- Powered Ultrasound Devices

Michiel A. P. Pertijs, Mingliang Tan, Nuriel N. M. Rozsa, Imad Bellouki et al.
2 citations
IEEE Journal of Solid-State Circuits
Ultrasound and Hyperthermia Applications
0.37
article

An Amplitude-Programmable Energy-Recycling High-Voltage Resonant Pulser for Battery- Powered Ultrasound Devices

Michiel A. P. Pertijs, Mingliang Tan, Nuriel N. M. Rozsa, Imad Bellouki, Zhao Chen, Zu‐Yao Chang
article en
2 citations

Abstract

This article presents an application-specific integrated circuit (ASIC) for battery-powered ultrasound (US) devices. The ASIC implements a novel energy-efficient high-voltage (HV) pulser that generates HV transmit (TX) pulses directly from a low-voltage (LV) battery supply. By means of a single off-chip inductor, energy is supplied to a US transducer in a resonant fashion, directly generating half-period sinusoidal HV pulses on the transducer, while consuming substantially less energy than a conventional class-D pulser. By recycling residual reactive energy from the transducer back to the input, the energy consumption is further reduced by more than 50%. The autocalibration techniques are leveraged to deal with tolerances of the inductor, transducer, and battery supply and thus maximize the energy efficiency. A prototype chip was fabricated in TSMC 0.18-<inline-formula xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink"> <tex-math notation="LaTeX">$\mu $ </tex-math></inline-formula>m HV BCD technology and used to drive external 120-pF capacitive micromachined US transducers (CMUTs) with a center frequency of approximately 2.5 MHz. Electrical measurements show that the prototype can generate pulses with a peak amplitude between 10 and 30 V accurate to within ±1 V. Acoustic measurements demonstrate successful ultrasonic pulse transmission and pulse-echo measurements. The prototype reaches a peak efficiency of 0.23 fCV2, which is the highest reported to date for HV pulsers targeting US imaging.

IEEE Journal of Solid-State CircuitsVol. 60(6)
Delft University of Technology (NL)
Affordable and clean energy
Openalex Percentile: Top 42%
Ultrasound and Hyperthermia Applications
0.37
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.