GTac-SE: A Low-Density Intention Sensing for Forearm Exoskeleton With Shear Force Sensors

This study explores the use of GTac-SE shear force sensors for detecting forearm rotation in upper-limb assistive exoskeletons for patients with arm weakness. The purpose is to evaluate these sensors as a reliable alternative to conventional methods, like electromyography (EMG) and inertial measurement units (IMUs), in real-world scenarios. Methods involved two experiments comparing GTac-SE sensors with EMG and IMUs. The first experiment assessed their ability to detect small forearm rotations mimicking weak patient motions, while the second evaluated sensor robustness during proximal joint (elbow) motions. Results showed that EMG struggled to detect subtle forearm rotations due to signal overlap between movement and rest, whereas IMUs and shear force sensors performed better. EMG was also sensitive to proximal joint movements, making intention detection difficult. While IMU gyroscopes displayed clear motion patterns, they were affected by other body movements, leading to unreliable joint angle estimations. In contrast, GTac-SE sensors, housed in a brace designed to isolate forearm rotation, outperformed both EMG and IMUs. GTac-SE signals were consistent across subjects with minimal interference during elbow flexion/extension. Additionally, GTac-SE required only simple moving-average filtering, while EMG and IMU signals needed complex filtering. In conclusion, GTac-SE shear force sensors provide a direct, reliable, and practical solution for intention sensing in daily-use exoskeletons. They outperform EMG and IMUs, particularly in scenarios where multijoint tasks are required for patients with weak arms.

Authors

Institutions

Publication Details

Journal
IEEE Sensors Journal
Published
2024-11-19
DOI
https://doi.org/10.1109/jsen.2024.3496904
Citations
2
Primary Topic
Muscle activation and electromyography studies
Type
article
Field-Weighted Citation Impact
0.37
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

GTac-SE: A Low-Density Intention Sensing for Forearm Exoskeleton With Shear Force Sensors

Thomas M. Kwok, Zeyu Lu, Haoyong Yu, Hilary HY Cheng
2 citations
IEEE Sensors Journal
Muscle activation and electromyography studies
0.37
article

GTac-SE: A Low-Density Intention Sensing for Forearm Exoskeleton With Shear Force Sensors

Thomas M. Kwok, Zeyu Lu, Haoyong Yu, Hilary HY Cheng
article en
2 citations

Abstract

This study explores the use of GTac-SE shear force sensors for detecting forearm rotation in upper-limb assistive exoskeletons for patients with arm weakness. The purpose is to evaluate these sensors as a reliable alternative to conventional methods, like electromyography (EMG) and inertial measurement units (IMUs), in real-world scenarios. Methods involved two experiments comparing GTac-SE sensors with EMG and IMUs. The first experiment assessed their ability to detect small forearm rotations mimicking weak patient motions, while the second evaluated sensor robustness during proximal joint (elbow) motions. Results showed that EMG struggled to detect subtle forearm rotations due to signal overlap between movement and rest, whereas IMUs and shear force sensors performed better. EMG was also sensitive to proximal joint movements, making intention detection difficult. While IMU gyroscopes displayed clear motion patterns, they were affected by other body movements, leading to unreliable joint angle estimations. In contrast, GTac-SE sensors, housed in a brace designed to isolate forearm rotation, outperformed both EMG and IMUs. GTac-SE signals were consistent across subjects with minimal interference during elbow flexion/extension. Additionally, GTac-SE required only simple moving-average filtering, while EMG and IMU signals needed complex filtering. In conclusion, GTac-SE shear force sensors provide a direct, reliable, and practical solution for intention sensing in daily-use exoskeletons. They outperform EMG and IMUs, particularly in scenarios where multijoint tasks are required for patients with weak arms.

IEEE Sensors JournalVol. 25(2)
National University of Singapore (SG)
Openalex Percentile: Top 42%
Muscle activation and electromyography studies
0.37
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.