Evaluation of plasma activated liquids for the elimination of mixed species biofilms within endoscopic working channels

The use of reusable flexible endoscopes has increased dramatically over the past decade, however despite improvements in endoscope reprocessing, the continued emergence of endoscopy-associated outbreaks as a result of multi-drug resistant bacteria has highlighted the need for a new approach to disinfection. Here, the use of plasma activated liquids (PALs) for the elimination of mixed species biofilm contamination within the working channels of endoscopes was evaluated. Cold atmospheric pressure plasma was used to chemically activate water and a commercially available pH buffered peracetic acid to create PALs. Polytetrafluoroethylene endoscope surrogate test pieces were contaminated with clinically relevant mixed species biofilms. The efficacy of PALs for the decontamination of narrow lumens was compared against the commercial disinfectant. Plasma activation was found to increase the antibiofilm capabilities of pH buffered peracetic acid by introducing reactive chemical species into the solution. Disinfection of endoscopic test pieces with plasma activated disinfectant (PAD) resulted in a 7.30 log<sub>10</sub> reduction of biofilm contamination in 5 min, surpassing the 4.39 log<sub>10</sub> reduction observed with the currently used endoscope disinfection method. PAD also resulted in reduced regrowth and recolonization of the surface of the endoscopic test pieces. Minimal changes to the surface morphology and composition were observed following exposure to PAD in comparison to the commercial disinfectant, suggesting the developed approach is no more aggressive than current disinfection approaches.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2024-11-19
DOI
https://doi.org/10.1038/s41598-024-79276-4
Citations
5
Primary Topic
Medical Device Sterilization and Disinfection
Type
article
Field-Weighted Citation Impact
1.72

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of plasma activated liquids for the elimination of mixed species biofilms within endoscopic working channels

Thorsten Juergens, Martina Modic, Vasyl Shvalya, Sascha Eschborn et al.
5 citations
Scientific Reports
Medical Device Sterilization and Disinfection
1.72
article

Evaluation of plasma activated liquids for the elimination of mixed species biofilms within endoscopic working channels

Thorsten Juergens, Martina Modic, Vasyl Shvalya, Sascha Eschborn, James L. Walsh, Naomi Northage, Malcolm J. Horsburgh
article en
5 citations

Abstract

The use of reusable flexible endoscopes has increased dramatically over the past decade, however despite improvements in endoscope reprocessing, the continued emergence of endoscopy-associated outbreaks as a result of multi-drug resistant bacteria has highlighted the need for a new approach to disinfection. Here, the use of plasma activated liquids (PALs) for the elimination of mixed species biofilm contamination within the working channels of endoscopes was evaluated. Cold atmospheric pressure plasma was used to chemically activate water and a commercially available pH buffered peracetic acid to create PALs. Polytetrafluoroethylene endoscope surrogate test pieces were contaminated with clinically relevant mixed species biofilms. The efficacy of PALs for the decontamination of narrow lumens was compared against the commercial disinfectant. Plasma activation was found to increase the antibiofilm capabilities of pH buffered peracetic acid by introducing reactive chemical species into the solution. Disinfection of endoscopic test pieces with plasma activated disinfectant (PAD) resulted in a 7.30 log<sub>10</sub> reduction of biofilm contamination in 5 min, surpassing the 4.39 log<sub>10</sub> reduction observed with the currently used endoscope disinfection method. PAD also resulted in reduced regrowth and recolonization of the surface of the endoscopic test pieces. Minimal changes to the surface morphology and composition were observed following exposure to PAD in comparison to the commercial disinfectant, suggesting the developed approach is no more aggressive than current disinfection approaches.

Scientific ReportsVol. 14(1)
University of Liverpool (GB), Jožef Stefan Institute (SI), Olympus (Germany) (DE), University of York (GB)
Javna Agencija za Raziskovalno Dejavnost RS, Engineering and Physical Sciences Research Council
Openalex Percentile: Top 15%
Medical Device Sterilization and Disinfection
1.72
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.