Estimation of the Current Uncertainty in the Dielectric Shoe Test According to the ISO/IEC 17025 Standard in the High Voltage Laboratory LABAV of the Escuela Politécnica Nacional

The High Voltage Laboratory (LABAV) at the Escuela Politécnica Nacional conducts dielectric tests on safety shoes in accordance with the ASTM F2412-18 standard. Additionally, as per the NTE INEN ISO 17025 standard, the laboratory must estimate the uncertainty of its measurements. Despite the scarcity of examples in the existing literature, this work provides a real-world example to assist other laboratories in replicating the uncertainty estimation process. In this article, we systematically present the calculation of leakage current uncertainty in shoes using both the traditional “Guide to the Expression of Uncertainty in Measurement” (GUM) method and the Monte Carlo method (MCM) for validation. The results from both approaches yield a similar uncertainty value of u = 0.0733 mA. Finally, we highlight the advantages that the MCM method offers in this context.

Authors

Institutions

Publication Details

Published
2024-11-07
DOI
https://doi.org/10.3390/engproc2024077021
Primary Topic
Advanced Sensor Technologies Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Estimation of the Current Uncertainty in the Dielectric Shoe Test According to the ISO/IEC 17025 Standard in the High Voltage Laboratory LABAV of the Escuela Politécnica Nacional

Edison Novoa, Jorge Medina, Alex Valenzuela, Diego Ontiveros Ramírez et al.
Advanced Sensor Technologies Research
article

Estimation of the Current Uncertainty in the Dielectric Shoe Test According to the ISO/IEC 17025 Standard in the High Voltage Laboratory LABAV of the Escuela Politécnica Nacional

Edison Novoa, Jorge Medina, Alex Valenzuela, Diego Ontiveros Ramírez, Juan Ramírez, William Chamorro, Victoria Paca
article en

Abstract

The High Voltage Laboratory (LABAV) at the Escuela Politécnica Nacional conducts dielectric tests on safety shoes in accordance with the ASTM F2412-18 standard. Additionally, as per the NTE INEN ISO 17025 standard, the laboratory must estimate the uncertainty of its measurements. Despite the scarcity of examples in the existing literature, this work provides a real-world example to assist other laboratories in replicating the uncertainty estimation process. In this article, we systematically present the calculation of leakage current uncertainty in shoes using both the traditional “Guide to the Expression of Uncertainty in Measurement” (GUM) method and the Monte Carlo method (MCM) for validation. The results from both approaches yield a similar uncertainty value of u = 0.0733 mA. Finally, we highlight the advantages that the MCM method offers in this context.

Universidad Técnica de Cotopaxi (EC), National Polytechnic School (EC)
Openalex Percentile: Top 81%
Advanced Sensor Technologies Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Estimation of the Current Uncertainty in the Dielectric Shoe Test According to the ISO/IEC 17025 Standard in the High Voltage Laboratory LABAV of the Escuela Politécnica Nacional — Edison Novoa, Jorge Medina, et al. · (2024) | TGRS Research Map | TGRS