Liposuction with real-time monitoring using Langevin transducer and imaging array

We constructed a liposuction therapeutic system using a Langevin transducer operating at 38 kHz, paired with an 8.2 MHz imaging array. This system generated high-power ultrasonic energy, which was applied to chicken breast tissue to validate its therapeutic potential. Simultaneously, the acoustic cavitation phenomena induced by the high-power ultrasound were recorded by an 8.2 MHz high-frequency imaging transducer and a hydrophone. With a voltage peak-to-peak (V PP ) amplitude of 150 V, the Langevin transducer emitted a peak-to-peak pressure of 31.56 kPa. This resulted in a broadband noise up to 200 kHz, indicative of transient cavitation. The imaging array visualized hyperechoic images resulting from cavitation, and the luminance of these images was evaluated as a function of electrical voltage. A comparative analysis between 50 V and 100 V indicated luminance values of 146 and 178, respectively, suggesting that higher electrical amplitudes correlate with increased bubble activity, leading to a wider noise band and brighter images. Additionally, the temperature rise induced by the Langevin transducer was investigated using a thermocouple , revealing that mechanical and thermal effects could occur simultaneously. A liposuction system with real-time cavitation monitoring will improve its efficacy through guided treatment. • Acoustic cavitation driven by 38 kHz Langevin transducer is shown to be monitored with linear array and spectral noise analysis. • Broadband noise up to 200 kHz and high echogenicity in 2D image are shown to exhibit the presence of transient cavitation. • The results suggest that liposuction surgery can be monitored in real-time with improved efficiency and reduced operation time.

Authors

Institutions

Publication Details

Journal
Sensors and Actuators A Physical
Published
2024-11-18
DOI
https://doi.org/10.1016/j.sna.2024.116063
Citations
1
Primary Topic
Body Contouring and Surgery
Type
article
Field-Weighted Citation Impact
0.57

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Liposuction with real-time monitoring using Langevin transducer and imaging array

Hyung Ham Kim, Yeonggeun Kim, Jungwoo Lee, Jinhyuk Kim
1 citations
Sensors and Actuators A Physical
Body Contouring and Surgery
0.57
article

Liposuction with real-time monitoring using Langevin transducer and imaging array

Hyung Ham Kim, Yeonggeun Kim, Jungwoo Lee, Jinhyuk Kim
article en
1 citations

Abstract

We constructed a liposuction therapeutic system using a Langevin transducer operating at 38 kHz, paired with an 8.2 MHz imaging array. This system generated high-power ultrasonic energy, which was applied to chicken breast tissue to validate its therapeutic potential. Simultaneously, the acoustic cavitation phenomena induced by the high-power ultrasound were recorded by an 8.2 MHz high-frequency imaging transducer and a hydrophone. With a voltage peak-to-peak (V PP ) amplitude of 150 V, the Langevin transducer emitted a peak-to-peak pressure of 31.56 kPa. This resulted in a broadband noise up to 200 kHz, indicative of transient cavitation. The imaging array visualized hyperechoic images resulting from cavitation, and the luminance of these images was evaluated as a function of electrical voltage. A comparative analysis between 50 V and 100 V indicated luminance values of 146 and 178, respectively, suggesting that higher electrical amplitudes correlate with increased bubble activity, leading to a wider noise band and brighter images. Additionally, the temperature rise induced by the Langevin transducer was investigated using a thermocouple , revealing that mechanical and thermal effects could occur simultaneously. A liposuction system with real-time cavitation monitoring will improve its efficacy through guided treatment. • Acoustic cavitation driven by 38 kHz Langevin transducer is shown to be monitored with linear array and spectral noise analysis. • Broadband noise up to 200 kHz and high echogenicity in 2D image are shown to exhibit the presence of transient cavitation. • The results suggest that liposuction surgery can be monitored in real-time with improved efficiency and reduced operation time.

Sensors and Actuators A PhysicalVol. 380
Pohang University of Science and Technology (KR), Kwangwoon University (KR)
Korea Medical Device Development Fund, Kwangwoon University, Ministry of Education, Ministry of Science and ICT, South Korea
Openalex Percentile: Top 27%
Body Contouring and Surgery
0.57
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.