On the impact of improved dose calculation accuracy in clinical treatment planning for superficial high-dose-rate brachytherapy of extensive scalp lesions

TG-43-based dose calculations disregard tissue heterogeneities and finite scatter conditions, prompting the introduction of model-based dose calculation algorithms (MBDCAs) to improve accuracy in high-dose-rate (HDR) brachytherapy. This study evaluated the effectiveness of MBDCAs over TG-43 in HDR <sup>192</sup>Ir brachytherapy of extended scalp lesions. Treatment planning dose calculations were compared with Monte Carlo (MC) data. TG-43 exhibited a dose overestimation ranging from 10% to 23% as the distance from the implant increased, while a better agreement from 2% to 6% was observed between the MBDCA and MC, supporting the adoption of MBDCAs for dose calculations in broad scalp lesions.

Authors

Institutions

Publication Details

Journal
Physics and Imaging in Radiation Oncology
Published
2024-10-01
DOI
https://doi.org/10.1016/j.phro.2024.100673
Citations
3
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
1.82
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

On the impact of improved dose calculation accuracy in clinical treatment planning for superficial high-dose-rate brachytherapy of extensive scalp lesions

Panagiotis Papagiannis, Vasiliki Peppa, Mark Gainey, Tanja Sprave et al.
3 citations
Physics and Imaging in Radiation Oncology
Advanced Radiotherapy Techniques
1.82
article

On the impact of improved dose calculation accuracy in clinical treatment planning for superficial high-dose-rate brachytherapy of extensive scalp lesions

Panagiotis Papagiannis, Vasiliki Peppa, Mark Gainey, Tanja Sprave, Dimos Baltas, M. Kollefrath, Giulio Rossi
article en
3 citations

Abstract

TG-43-based dose calculations disregard tissue heterogeneities and finite scatter conditions, prompting the introduction of model-based dose calculation algorithms (MBDCAs) to improve accuracy in high-dose-rate (HDR) brachytherapy. This study evaluated the effectiveness of MBDCAs over TG-43 in HDR <sup>192</sup>Ir brachytherapy of extended scalp lesions. Treatment planning dose calculations were compared with Monte Carlo (MC) data. TG-43 exhibited a dose overestimation ranging from 10% to 23% as the distance from the implant increased, while a better agreement from 2% to 6% was observed between the MBDCA and MC, supporting the adoption of MBDCAs for dose calculations in broad scalp lesions.

Physics and Imaging in Radiation OncologyVol. 32
National and Kapodistrian University of Athens (GR), University Medical Center Freiburg (DE)
Good health and well-being
Openalex Percentile: Top 15%
Advanced Radiotherapy Techniques
1.82
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.