Allowable total error in CD34 cell analysis by flow cytometry based on state of the art using Spanish EQAS data
Technical advancements contribute to the decrease TE over time. The TE of CD34 cell FC counts is measure-dependent, making it responsive to precision enhancement strategies. The TE measured by EQAS in this study may serve as a quality specification for implementing ISO 15189 standards in clinical laboratories for CD34 cell enumeration.