Strategies for Using Postcolumn Infusion of Standards to Correct for Matrix Effect in LC-MS-Based Quantitative Metabolomics

The matrix effect limits the accuracy of quantitation of the otherwise popular metabolomics technique liquid chromatography coupled to mass spectrometry (LC-MS). The gold standard to correct for this phenomenon, whereby compounds coeluting with the analyte of interest cause ionization enhancement or suppression, is to quantify an analyte based on the peak area ratio with an isotopologue added to the sample as an internal standard. However, these stable isotopes are expensive and sometimes unavailable. Here, we describe an alternative approach: matrix effect correction and quantifying analytes using a signal ratio with a postcolumn infused standard (PCIS). Using an LC-MS/MS method for eight endocannabinoids and related metabolites in plasma, we provide strategies to select, optimize, and evaluate PCIS candidates. Based on seven characteristics, the structural endocannabinoid analogue arachidonoyl-2'-fluoroethylamide was selected as a PCIS. Three methods to evaluate the PCIS correction vs no correction showed that PCIS correction improved values for the matrix effect, precision, and dilutional linearity of at least six of the analytes to within acceptable ranges. PCIS correction also resulted in parallelization of calibration curves in plasma and neat solution, for six of eight analytes even with higher accuracy than peak area ratio correction with their stable isotope labeled internal standard, i.e., the gold standard. This enables quantification based on neat solutions, which is a significant step toward absolute quantification. We conclude that PCIS has great, but so far underappreciated, potential in accurate LC-MS quantification.

Authors

Institutions

Publication Details

Journal
Journal of the American Society for Mass Spectrometry
Published
2024-11-15
DOI
https://doi.org/10.1021/jasms.4c00408
Citations
9
Primary Topic
Metabolomics and Mass Spectrometry Studies
Type
article
Field-Weighted Citation Impact
2.10

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Strategies for Using Postcolumn Infusion of Standards to Correct for Matrix Effect in LC-MS-Based Quantitative Metabolomics

Anne-Charlotte Dubbelman, Roel C.H. Vermeulen, Amy C. Harms, Thomas Hankemeier et al.
9 citations
Journal of the American Society for Mass Spectrometry
Metabolomics and Mass Spectrometry Studies
2.10
article

Strategies for Using Postcolumn Infusion of Standards to Correct for Matrix Effect in LC-MS-Based Quantitative Metabolomics

Anne-Charlotte Dubbelman, Roel C.H. Vermeulen, Amy C. Harms, Thomas Hankemeier, Michael Van Vliet, Bo van Wieringen, L. Arias
article en
9 citations

Abstract

The matrix effect limits the accuracy of quantitation of the otherwise popular metabolomics technique liquid chromatography coupled to mass spectrometry (LC-MS). The gold standard to correct for this phenomenon, whereby compounds coeluting with the analyte of interest cause ionization enhancement or suppression, is to quantify an analyte based on the peak area ratio with an isotopologue added to the sample as an internal standard. However, these stable isotopes are expensive and sometimes unavailable. Here, we describe an alternative approach: matrix effect correction and quantifying analytes using a signal ratio with a postcolumn infused standard (PCIS). Using an LC-MS/MS method for eight endocannabinoids and related metabolites in plasma, we provide strategies to select, optimize, and evaluate PCIS candidates. Based on seven characteristics, the structural endocannabinoid analogue arachidonoyl-2'-fluoroethylamide was selected as a PCIS. Three methods to evaluate the PCIS correction vs no correction showed that PCIS correction improved values for the matrix effect, precision, and dilutional linearity of at least six of the analytes to within acceptable ranges. PCIS correction also resulted in parallelization of calibration curves in plasma and neat solution, for six of eight analytes even with higher accuracy than peak area ratio correction with their stable isotope labeled internal standard, i.e., the gold standard. This enables quantification based on neat solutions, which is a significant step toward absolute quantification. We conclude that PCIS has great, but so far underappreciated, potential in accurate LC-MS quantification.

Journal of the American Society for Mass SpectrometryVol. 35(12)
Leiden University (NL), Utrecht University (NL), Centre for Human Drug Research (NL)
Nederlandse Organisatie voor Wetenschappelijk Onderzoek, Medical Delta
Zero hunger
Openalex Percentile: Top 12%
Metabolomics and Mass Spectrometry Studies
2.10
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.