AAPM task group report 314: Fault recovery in external beam radiation therapy

Task Group (TG) 314 of the American Association of Physicists in Medicine (AAPM) was charged to develop guidance for recovering from fault states in radiation therapy, specifically regarding the delivery of photon or electron beams using a linear accelerator (linac) including ancillary systems. The fault conditions addressed may involve software, hardware, or a combination of causes. The report provides detailed recommendations for the proactive steps to be taken before a fault, the actions to be taken at the time of a fault, and the safety steps before returning a linac to clinical service, as well as the activities that device manufacturers and standard organizations can do to prevent and resolve the faults. A user-maintained log of prior faults; establishment of remote access by the vendor; and user training in emergency gantry, couch, and door motions are all useful proactive steps. At the moment of downtime and after ensuring the safety of the patient, the report stresses the importance of capturing fault information, prompt contact with the service engineer after the initial assessment, and considerations for communicating the estimated duration before the linac is returned to service. The medical physicist has a critical responsibility to assess the impact of the fault on patient care. Before resuming clinical use, the medical physicist must both determine the level of testing required to ensure safe operation of the linac and ensure any partially or totally delivered treatments have been correctly saved for accurate completion of the treatment fraction. The report stresses the roles of the radiation therapist, medical physicist, and service engineer to efficiently and safely address linac downtime. The appendices contain a description of the efforts of several organizations regarding linac safety: Integrating the Healthcare Enterprise-Radiation Oncology, International Standards Organization/International Electrotechnical Commission, Radiation Oncology Safety Stakeholder Initiative, and the AAPM Vendor Relations and Product Usability Subcommittee. Disclaimer: The recommendations of this TG should not be used to establish regulations. These recommendations are guidelines for Qualified Medical Physicists and others to use and appropriately interpret for their institution and clinical setting. Each institution may have site-specific or state-mandated needs and requirements which may modify their usage of these recommendations.

Authors

Institutions

Publication Details

Journal
Medical Physics
Published
2024-11-15
DOI
https://doi.org/10.1002/mp.17502
Citations
6
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
3.65
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

AAPM task group report 314: Fault recovery in external beam radiation therapy

Scott Hadley, Sugata Tripathi, R Siochi, C Buckey et al.
6 citations
Medical Physics
Advanced Radiotherapy Techniques
3.65
article

AAPM task group report 314: Fault recovery in external beam radiation therapy

Scott Hadley, Sugata Tripathi, R Siochi, C Buckey, Geoffrey S. Ibbott, D Perrin, Steven Sutlief, Choi Bo Young, Ian Robertson, Bruce Curran, Cristina Negrut, Susan Koehl, Jie Shi
article en
6 citations

Abstract

Task Group (TG) 314 of the American Association of Physicists in Medicine (AAPM) was charged to develop guidance for recovering from fault states in radiation therapy, specifically regarding the delivery of photon or electron beams using a linear accelerator (linac) including ancillary systems. The fault conditions addressed may involve software, hardware, or a combination of causes. The report provides detailed recommendations for the proactive steps to be taken before a fault, the actions to be taken at the time of a fault, and the safety steps before returning a linac to clinical service, as well as the activities that device manufacturers and standard organizations can do to prevent and resolve the faults. A user-maintained log of prior faults; establishment of remote access by the vendor; and user training in emergency gantry, couch, and door motions are all useful proactive steps. At the moment of downtime and after ensuring the safety of the patient, the report stresses the importance of capturing fault information, prompt contact with the service engineer after the initial assessment, and considerations for communicating the estimated duration before the linac is returned to service. The medical physicist has a critical responsibility to assess the impact of the fault on patient care. Before resuming clinical use, the medical physicist must both determine the level of testing required to ensure safe operation of the linac and ensure any partially or totally delivered treatments have been correctly saved for accurate completion of the treatment fraction. The report stresses the roles of the radiation therapist, medical physicist, and service engineer to efficiently and safely address linac downtime. The appendices contain a description of the efforts of several organizations regarding linac safety: Integrating the Healthcare Enterprise-Radiation Oncology, International Standards Organization/International Electrotechnical Commission, Radiation Oncology Safety Stakeholder Initiative, and the AAPM Vendor Relations and Product Usability Subcommittee. Disclaimer: The recommendations of this TG should not be used to establish regulations. These recommendations are guidelines for Qualified Medical Physicists and others to use and appropriately interpret for their institution and clinical setting. Each institution may have site-specific or state-mandated needs and requirements which may modify their usage of these recommendations.

Medical PhysicsVol. 52(1)
West Virginia University (US), The University of Texas MD Anderson Cancer Center (US), Virginia Commonwealth University (US), University of Michigan (US), WinnMed (US), Accuray (United States) (US), National University (US), Sharp HealthCare Foundation (US), Sun Nuclear (United States) (US), Marshfield Clinic (US), Elekta (United Kingdom) (GB), Mayo Clinic in Florida (US), Mary Bird Perkins Cancer Center (US), Siemens Healthcare (United States) (US)
Partnerships for the goals
Openalex Percentile: Top 7%
Advanced Radiotherapy Techniques
3.65
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.