Abstract Su1011: Effect of Constriction Rates on Indicators of Hemodynamics in Circumferential Thoracoabdominal Cardiopulmonary Resuscitation

Introduction: Data from human case series suggest that the optimal compression rate for sternal piston type CPR may be in the range of 130 compressions/min. Existing mechanical devices have compression rates of between 80 and 100 per minute. Circumferential thoracoabdominal CPR ( CTA-CPR ) may have enhanced hemodynamic effects and a superior safety profile, but the optimal constriction rate is currently unknown. Hypothesis: Constriction rates are positively correlated with indicators of hemodynamics in CTA-CPR. Methods: Ventricular fibrillation was induced in female Yorkshire swine (n = 9, ~30kg). CPR was provided by a prototype pneumatic CTA-CPR system. Constriction rates were varied in 45s epochs among 80, 100 and 120 per minute. Aortic ( Ao ) and right atrial ( RA ) pressures were measured using micromanometer catheters. Coronary perfusion pressure ( CPP ) was derived as the difference between the Ao and RA pressures. The change in CPP ( Δ-CPP ) from the beginning to the end of each treatment epoch was the principal indicator of hemodynamics. Data from all experiments were grouped by treatment type and subjected to Analysis of Variance. Results: Final Ao pressures for 80, 100, and 120 constrictions per minute were 27, 31, and 28 mmHg, respectively. Final RA pressures were 6, 7, and 7 mmHg. This yields CPPs of 20, 24, and 21 mmHg. The Δ-CPPs trended upward 5.3, 5.7, and 6.0, respectively (p < 0.10). Conclusion: Increasing constriction rates were associated with improved indicators of hemodynamics. The optimal rate may be similar to that seen in sternal piston CPR, and considerably higher than available with commercial automated devices.

Authors

Institutions

Publication Details

Journal
Circulation
Published
2024-11-12
DOI
https://doi.org/10.1161/circ.150.suppl_1.su1011
Primary Topic
Cardiac Arrest and Resuscitation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract Su1011: Effect of Constriction Rates on Indicators of Hemodynamics in Circumferential Thoracoabdominal Cardiopulmonary Resuscitation

Alexandre Dufresne, David Gaddy, Karen L Moodie, Daniel Davis et al.
Circulation
Cardiac Arrest and Resuscitation
article

Abstract Su1011: Effect of Constriction Rates on Indicators of Hemodynamics in Circumferential Thoracoabdominal Cardiopulmonary Resuscitation

Alexandre Dufresne, David Gaddy, Karen L Moodie, Daniel Davis, Norman A. Paradis, Christine S. Couturier, Simon F. Dufresne, Christopher E. Sims, Timothy J. Mader, Aidan Paradis
article en

Abstract

Introduction: Data from human case series suggest that the optimal compression rate for sternal piston type CPR may be in the range of 130 compressions/min. Existing mechanical devices have compression rates of between 80 and 100 per minute. Circumferential thoracoabdominal CPR ( CTA-CPR ) may have enhanced hemodynamic effects and a superior safety profile, but the optimal constriction rate is currently unknown. Hypothesis: Constriction rates are positively correlated with indicators of hemodynamics in CTA-CPR. Methods: Ventricular fibrillation was induced in female Yorkshire swine (n = 9, ~30kg). CPR was provided by a prototype pneumatic CTA-CPR system. Constriction rates were varied in 45s epochs among 80, 100 and 120 per minute. Aortic ( Ao ) and right atrial ( RA ) pressures were measured using micromanometer catheters. Coronary perfusion pressure ( CPP ) was derived as the difference between the Ao and RA pressures. The change in CPP ( Δ-CPP ) from the beginning to the end of each treatment epoch was the principal indicator of hemodynamics. Data from all experiments were grouped by treatment type and subjected to Analysis of Variance. Results: Final Ao pressures for 80, 100, and 120 constrictions per minute were 27, 31, and 28 mmHg, respectively. Final RA pressures were 6, 7, and 7 mmHg. This yields CPPs of 20, 24, and 21 mmHg. The Δ-CPPs trended upward 5.3, 5.7, and 6.0, respectively (p < 0.10). Conclusion: Increasing constriction rates were associated with improved indicators of hemodynamics. The optimal rate may be similar to that seen in sternal piston CPR, and considerably higher than available with commercial automated devices.

CirculationVol. 150(Suppl_1)
Baystate Medical Center (US), University of Massachusetts Chan Medical School (US), University of California San Diego (US), Baystate Health (US)
Good health and well-being
Openalex Percentile: Top 63%
Cardiac Arrest and Resuscitation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.