Abstract 4139677: A rare case of ventriculobronchial fistula caused by an epicardial defibrillator patch

Background: Epicardial patch defibrillators (EPDs) were commonly implanted in the 1990s for secondary prevention of sudden cardiac death. Despite being largely obsolete, some EPDs remain in patients and can cause late complications. This case highlights such a scenario. Case: A 75-year-old female with a history of cardiac arrest 30 years ago presented with shortness of breath and left leg swelling. She had idiopathic ventricular fibrillation in 1992, treated with an EPD (Picture 1A), later replaced by a transvenous ICD. She was diagnosed with left femoral deep venous thrombosis and bilateral pulmonary embolism and started on therapeutic anticoagulation. She experienced massive hemoptysis, leading to respiratory and cardiac arrest, but was resuscitated. A CT angiogram showed no active bleeding but noted the EPD was close to the lower lobe of the left lung. Upper gastrointestinal endoscopy showed no active bleeding. Bronchoscopy revealed clots in the left lower lobe. Despite empiric bronchial artery embolization, hemoptysis persisted. A repeat bronchoscopy showed reaccumulation of clots. A repeat CT angiogram indicated contrast extension into the ventricular myocardium near the EPD but no lung spillage suggestive of pseudoaneurysm (Picture 1B). She was transferred to our institution for further evaluation. A coronary angiogram revealed normal arteries, while a left ventriculogram revealed contrast extravasation through the lateral wall (Picture 1C). An urgent surgery revealed two EPDs: anterior and lateral. The lateral EPD had eroded into the ventricle, creating a ventriculo bronchial fistula (Picture 1D). The lateral EPD was removed (Picture 1E); the anterior EPD was left in place (Picture 1F). The ventricular defect was debrided and repaired. Postoperative recovery was successful, with no further hemoptysis, and the patient was discharged to a rehabilitation facility. Discussion: EPDs are no longer used due to high failure rates and complications like scar formation and constrictive pericarditis. However, patients with EPDs from the 1990s may present with delayed complications. Accurate diagnosis using multimodality imaging and early management is crucial to reduce morbidity and mortality.

Authors

Institutions

Publication Details

Journal
Circulation
Published
2024-11-12
DOI
https://doi.org/10.1161/circ.150.suppl_1.4139677
Primary Topic
Cardiac Structural Anomalies and Repair
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract 4139677: A rare case of ventriculobronchial fistula caused by an epicardial defibrillator patch

JOHN WINDLE, Sai Vikram Alampoondi Venkataramanan
Circulation
Cardiac Structural Anomalies and Repair
article

Abstract 4139677: A rare case of ventriculobronchial fistula caused by an epicardial defibrillator patch

JOHN WINDLE, Sai Vikram Alampoondi Venkataramanan
article en

Abstract

Background: Epicardial patch defibrillators (EPDs) were commonly implanted in the 1990s for secondary prevention of sudden cardiac death. Despite being largely obsolete, some EPDs remain in patients and can cause late complications. This case highlights such a scenario. Case: A 75-year-old female with a history of cardiac arrest 30 years ago presented with shortness of breath and left leg swelling. She had idiopathic ventricular fibrillation in 1992, treated with an EPD (Picture 1A), later replaced by a transvenous ICD. She was diagnosed with left femoral deep venous thrombosis and bilateral pulmonary embolism and started on therapeutic anticoagulation. She experienced massive hemoptysis, leading to respiratory and cardiac arrest, but was resuscitated. A CT angiogram showed no active bleeding but noted the EPD was close to the lower lobe of the left lung. Upper gastrointestinal endoscopy showed no active bleeding. Bronchoscopy revealed clots in the left lower lobe. Despite empiric bronchial artery embolization, hemoptysis persisted. A repeat bronchoscopy showed reaccumulation of clots. A repeat CT angiogram indicated contrast extension into the ventricular myocardium near the EPD but no lung spillage suggestive of pseudoaneurysm (Picture 1B). She was transferred to our institution for further evaluation. A coronary angiogram revealed normal arteries, while a left ventriculogram revealed contrast extravasation through the lateral wall (Picture 1C). An urgent surgery revealed two EPDs: anterior and lateral. The lateral EPD had eroded into the ventricle, creating a ventriculo bronchial fistula (Picture 1D). The lateral EPD was removed (Picture 1E); the anterior EPD was left in place (Picture 1F). The ventricular defect was debrided and repaired. Postoperative recovery was successful, with no further hemoptysis, and the patient was discharged to a rehabilitation facility. Discussion: EPDs are no longer used due to high failure rates and complications like scar formation and constrictive pericarditis. However, patients with EPDs from the 1990s may present with delayed complications. Accurate diagnosis using multimodality imaging and early management is crucial to reduce morbidity and mortality.

CirculationVol. 150(Suppl_1)
University of Nebraska Medical Center (US)
Good health and well-being
Openalex Percentile: Top 61%
Cardiac Structural Anomalies and Repair
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.