Abstract 4140687: Characterization of the Near-Field Bipolar Voltage Substrate Using a Novel Automated Peak Frequency Algorithm in Patients Undergoing Catheter Ablation of Postinfarction Ventricular Tachycardia

Background: Ventricular tachycardia (VT) substrate characterization in patients with ischemic cardiomyopathy (ICM) utilizing conventional peak-to-peak bipolar voltage mapping is limited by preferential automatic annotation of large amplitude far-field electrograms (EGMs). Objective: To investigate the role of a novel peak frequency (PF) software that prioritizes annotation of the near-field high-frequency component of bipolar EGMs as compared to standard bipolar voltage mapping (Std-Bip) for postinfarction substrate characterization. Methods: We studied 23 patients (age 67.2±9.4 years, 95.7% male, LVEF 37.4±14.6%) undergoing catheter ablation of postinfarction VT (34.8% anterior infarction, 73.9% inferior infarction). High-density endocardial substrate maps were created in sinus rhythm (43.5%) or right ventricular pacing (56.5%) using a 16-electrode mapping catheter (HD grid, Abbott, IL). The peak-to-peak bipolar voltage of collected EGMs was analyzed in Std-Bip configuration and after applying the PF software. Results: A total of 105,815 Std-Bip and PF EGMs were analyzed. Overall, 30.6% Std-Bip EGMs had a voltage 0.5-1.5 mV and 39% a voltage <0.5 mV. After applying the PF software, 26.3% PF EGMs had a voltage 0.5-1.5 mV and 48.4% a voltage <0.5 mV (p<0.01 for comparison, Figure ). The PF software reclassified an average of 8.9% (range 3.1% to 21.9%) Std-Bip EGMs to a voltage <0.5 mV and an average 3.5% (range -7.9% to 13.5%) Std-Bip EGMs to a voltage 0.5-1.5 mV. This resulted in an average increase in LV surface area subtending low voltage (<1.5 mV) EGMs of 3.9±0.03%. Conclusions: Postinfarction substrate mapping with a PF annotation algorithm significantly improves identification of low voltage bipolar EGMs compared to Std-Bip mapping.

Authors

Institutions

Publication Details

Journal
Circulation
Published
2024-11-12
DOI
https://doi.org/10.1161/circ.150.suppl_1.4140687
Primary Topic
Cardiac Arrhythmias and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract 4140687: Characterization of the Near-Field Bipolar Voltage Substrate Using a Novel Automated Peak Frequency Algorithm in Patients Undergoing Catheter Ablation of Postinfarction Ventricular Tachycardia

Roy Chung, Arwa Younis, Mohamed Kanj, Shady Nakhla et al.
Circulation
Cardiac Arrhythmias and Treatments
article

Abstract 4140687: Characterization of the Near-Field Bipolar Voltage Substrate Using a Novel Automated Peak Frequency Algorithm in Patients Undergoing Catheter Ablation of Postinfarction Ventricular Tachycardia

Roy Chung, Arwa Younis, Mohamed Kanj, Shady Nakhla, Walid I. Saliba, Jakub Sroubek, Ayman A. Hussein, Pasquale Santangeli, Hiroshi Nakagawa, Mandeep Bhargava, Oussama M. Wazni, Koji Higuchi, Justin Lee, Monica Korzon, Manojna Nimmagadda
article en

Abstract

Background: Ventricular tachycardia (VT) substrate characterization in patients with ischemic cardiomyopathy (ICM) utilizing conventional peak-to-peak bipolar voltage mapping is limited by preferential automatic annotation of large amplitude far-field electrograms (EGMs). Objective: To investigate the role of a novel peak frequency (PF) software that prioritizes annotation of the near-field high-frequency component of bipolar EGMs as compared to standard bipolar voltage mapping (Std-Bip) for postinfarction substrate characterization. Methods: We studied 23 patients (age 67.2±9.4 years, 95.7% male, LVEF 37.4±14.6%) undergoing catheter ablation of postinfarction VT (34.8% anterior infarction, 73.9% inferior infarction). High-density endocardial substrate maps were created in sinus rhythm (43.5%) or right ventricular pacing (56.5%) using a 16-electrode mapping catheter (HD grid, Abbott, IL). The peak-to-peak bipolar voltage of collected EGMs was analyzed in Std-Bip configuration and after applying the PF software. Results: A total of 105,815 Std-Bip and PF EGMs were analyzed. Overall, 30.6% Std-Bip EGMs had a voltage 0.5-1.5 mV and 39% a voltage <0.5 mV. After applying the PF software, 26.3% PF EGMs had a voltage 0.5-1.5 mV and 48.4% a voltage <0.5 mV (p<0.01 for comparison, Figure ). The PF software reclassified an average of 8.9% (range 3.1% to 21.9%) Std-Bip EGMs to a voltage <0.5 mV and an average 3.5% (range -7.9% to 13.5%) Std-Bip EGMs to a voltage 0.5-1.5 mV. This resulted in an average increase in LV surface area subtending low voltage (<1.5 mV) EGMs of 3.9±0.03%. Conclusions: Postinfarction substrate mapping with a PF annotation algorithm significantly improves identification of low voltage bipolar EGMs compared to Std-Bip mapping.

CirculationVol. 150(Suppl_1)
Cleveland Clinic (US), Abbott (United States) (US)
Good health and well-being
Openalex Percentile: Top 62%
Cardiac Arrhythmias and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.