Abstract 4142224: Efficacy of Inferior Vena Cava-Pulmonary Artery Bypass for Right Ventricular Dysfunction under Left Ventricular Assist Device Support: An in vitro study using a biventricular circulation system

Introduction: Right heart failure under the support of a left ventricular assist device (LVAD) presents a life-threatening condition characterized by organ edema and limited LVAD support. Implementing an inferior vena cava-pulmonary artery (IVC-PA) bypass graft may mitigate these complications by reducing central venous pressure (CVP) and improving LVAD support efficiency. This study aimed to elucidate an adequate graft diameter for an IVC-PA bypass in terms of CVP reduction during LVAD-assisted circulation using an in vitro biventricular pulsatile circulatory system. Methods: We developed a biventricular pulsatile circulation system capable of providing circulatory assistance via an LVAD (centrifugal pump) connected to the apex of a left ventricular model (Fig). A representative condition of cardiogenic shock was produced by adjusting LV systolic pressure, aortic pressure, and CVP to 80 mmHg, 80/40 mmHg, and 7.5 mmHg, respectively. A right heart failure model was produced by adjusting LV systolic pressure, aortic pressure, and CVP to 45 mmHg, 70 mmHg, and 16 mmHg, respectively, under LVAD support at 1700 rpm. Then, an adequate IVC-PA bypass diameter was investigated in terms of reducing CVP and increasing bypass flow support among 4mm and 18 mm with 2mm interval. The diameters of PA and IVC, and systolic PA pressure were set to 19 mm, 19 mm, and 18 mmHg, respectively. Elastic LV and RV models were driven by pneumatic positive and negative pressures. Results: We found a linearly increasing trend of bypass flow and a decreasing trend of CVP with increasing bypass diameters from 4 to 12 mm. When applying the bypass graft over 12mm diameter, bypass flow and CVP plateaued at 3.4 L/min and 7.5 mmHg, respectively (Fig). Conclusion: Our sophisticated in vitro biventricular circulation study suggests that in scenarios of right heart failure under LVAD support, implementing a bypass graft from the IVC to the PA is effective for decreasing CVP and reducing right ventricular preload. This study indicates that the optimal bypass diameter for reducing CVP is 12 mm when the diameters of the IVC and PA are 19 mm. These findings encouraged us to evaluate the efficacy in in vivo clinical settings.

Authors

Institutions

Publication Details

Journal
Circulation
Published
2024-11-12
DOI
https://doi.org/10.1161/circ.150.suppl_1.4142224
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract 4142224: Efficacy of Inferior Vena Cava-Pulmonary Artery Bypass for Right Ventricular Dysfunction under Left Ventricular Assist Device Support: An in vitro study using a biventricular circulation system

Kiyotaka Iwasaki, Kojiro Furukawa, Kaoru Hattori, Jumpei Takada et al.
Circulation
Mechanical Circulatory Support Devices
article

Abstract 4142224: Efficacy of Inferior Vena Cava-Pulmonary Artery Bypass for Right Ventricular Dysfunction under Left Ventricular Assist Device Support: An in vitro study using a biventricular circulation system

Kiyotaka Iwasaki, Kojiro Furukawa, Kaoru Hattori, Jumpei Takada, Hitoshi Inafuku, Keiji Uchida
article en

Abstract

Introduction: Right heart failure under the support of a left ventricular assist device (LVAD) presents a life-threatening condition characterized by organ edema and limited LVAD support. Implementing an inferior vena cava-pulmonary artery (IVC-PA) bypass graft may mitigate these complications by reducing central venous pressure (CVP) and improving LVAD support efficiency. This study aimed to elucidate an adequate graft diameter for an IVC-PA bypass in terms of CVP reduction during LVAD-assisted circulation using an in vitro biventricular pulsatile circulatory system. Methods: We developed a biventricular pulsatile circulation system capable of providing circulatory assistance via an LVAD (centrifugal pump) connected to the apex of a left ventricular model (Fig). A representative condition of cardiogenic shock was produced by adjusting LV systolic pressure, aortic pressure, and CVP to 80 mmHg, 80/40 mmHg, and 7.5 mmHg, respectively. A right heart failure model was produced by adjusting LV systolic pressure, aortic pressure, and CVP to 45 mmHg, 70 mmHg, and 16 mmHg, respectively, under LVAD support at 1700 rpm. Then, an adequate IVC-PA bypass diameter was investigated in terms of reducing CVP and increasing bypass flow support among 4mm and 18 mm with 2mm interval. The diameters of PA and IVC, and systolic PA pressure were set to 19 mm, 19 mm, and 18 mmHg, respectively. Elastic LV and RV models were driven by pneumatic positive and negative pressures. Results: We found a linearly increasing trend of bypass flow and a decreasing trend of CVP with increasing bypass diameters from 4 to 12 mm. When applying the bypass graft over 12mm diameter, bypass flow and CVP plateaued at 3.4 L/min and 7.5 mmHg, respectively (Fig). Conclusion: Our sophisticated in vitro biventricular circulation study suggests that in scenarios of right heart failure under LVAD support, implementing a bypass graft from the IVC to the PA is effective for decreasing CVP and reducing right ventricular preload. This study indicates that the optimal bypass diameter for reducing CVP is 12 mm when the diameters of the IVC and PA are 19 mm. These findings encouraged us to evaluate the efficacy in in vivo clinical settings.

CirculationVol. 150(Suppl_1)
Waseda University (JP), University of the Ryukyus (JP)
Openalex Percentile: Top 81%
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.