Abstract 4143757: Adverse Outcomes with Cardiac Contractility Modulation: Insights From the MAUDE Registry

Background: Cardiac contractility modulation (CCM) is a device-based heart failure therapy that augments cardiac contractility. One CCM device – the Optimizer Smart Mini System [Impulse Dynamics, Marlton, NJ] – is approved by the Food and Drug Administration (FDA) for clinical use. Research Question: What are real-world and long-term outcomes of the Optimizer Smart Mini CCM system? Objective: We reviewed the United States FDA Manufacturer and User Facility Device Experience (MAUDE) database for adverse events and outcomes associated with the Optimizer Smart Mini System CCM. Methods: On June 3rd, 2024, we performed a MAUDE database search for all reports of CCM devices from January 2019 - when the first pre-market implantation was noted - to the present. Duplicates were removed, and outcomes were noted. Results: A total of 34 total reports were found. The majority of events were reported since 2023. 58.8% (n=20) of events were classified as device malfunctions. This occurred most often due to a computer software problem (35.3%, n=12), failure to charge (14.7%, n=5), and an adverse event without an identified problem (14.7%, n=5). Commonly, error code A09 - a charging error due to incorrect interpretation of a high current during a routine safety self-check - occurred in 38.2% (n=13) of all cases, even in patients without a reported software problem, and was solved by resetting the device in 47.1% (n=16) of cases. This is a known problem. Injury occurred in 35.3% (n=12) of cases due to implant failure (52.9%, n=18) as the most common patient problem. There was a 5.9% (n=2) rate of mortality. Conclusions: Real-world CCM implantation appears to have low device-related adverse risks but suffers from computer software issues that may be solved by resetting the device. Continued monitoring and accurate adverse event reporting are needed with increased use to assess long-term risks.

Authors

Institutions

Publication Details

Journal
Circulation
Published
2024-11-12
DOI
https://doi.org/10.1161/circ.150.suppl_1.4143757
Primary Topic
Cardiac Arrhythmias and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract 4143757: Adverse Outcomes with Cardiac Contractility Modulation: Insights From the MAUDE Registry

Dhanunjaya Lakkireddy, Donita Atkins, Naga Venkata K. Pothineni, Aashish Katapadi et al.
Circulation
Cardiac Arrhythmias and Treatments
article

Abstract 4143757: Adverse Outcomes with Cardiac Contractility Modulation: Insights From the MAUDE Registry

Dhanunjaya Lakkireddy, Donita Atkins, Naga Venkata K. Pothineni, Aashish Katapadi, Rajesh Kabra, Nikhila Chelikam, Douglas Darden, Rakesh Gopinathannair, Aashika Katapadi
article en

Abstract

Background: Cardiac contractility modulation (CCM) is a device-based heart failure therapy that augments cardiac contractility. One CCM device – the Optimizer Smart Mini System [Impulse Dynamics, Marlton, NJ] – is approved by the Food and Drug Administration (FDA) for clinical use. Research Question: What are real-world and long-term outcomes of the Optimizer Smart Mini CCM system? Objective: We reviewed the United States FDA Manufacturer and User Facility Device Experience (MAUDE) database for adverse events and outcomes associated with the Optimizer Smart Mini System CCM. Methods: On June 3rd, 2024, we performed a MAUDE database search for all reports of CCM devices from January 2019 - when the first pre-market implantation was noted - to the present. Duplicates were removed, and outcomes were noted. Results: A total of 34 total reports were found. The majority of events were reported since 2023. 58.8% (n=20) of events were classified as device malfunctions. This occurred most often due to a computer software problem (35.3%, n=12), failure to charge (14.7%, n=5), and an adverse event without an identified problem (14.7%, n=5). Commonly, error code A09 - a charging error due to incorrect interpretation of a high current during a routine safety self-check - occurred in 38.2% (n=13) of all cases, even in patients without a reported software problem, and was solved by resetting the device in 47.1% (n=16) of cases. This is a known problem. Injury occurred in 35.3% (n=12) of cases due to implant failure (52.9%, n=18) as the most common patient problem. There was a 5.9% (n=2) rate of mortality. Conclusions: Real-world CCM implantation appears to have low device-related adverse risks but suffers from computer software issues that may be solved by resetting the device. Continued monitoring and accurate adverse event reporting are needed with increased use to assess long-term risks.

CirculationVol. 150(Suppl_1)
Heart Rhythm Society (US), University of Aruba (AW)
Good health and well-being
Openalex Percentile: Top 62%
Cardiac Arrhythmias and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.