Atraumatic Failure of the DePuy Synthes Trochanteric Fixation Nail Advanced Fenestrated Cephalic Lag Screw
Case: Extracapsular proximal femur fractures are a common injury that can be reliably treated with cephalomedullary nailing. We present 2 cases of atraumatic DePuy Synthes Trochanteric Fixation Nail Advanced implant failure through cephalic lag screw fenestrations in the setting of intertrochanteric fracture. Conclusion: Fenestrated lag screws inserted without cement augmentation may be susceptible to fatigue failure. Patients with unstable fracture patterns and other risk factors for delayed healing should undergo close clinical follow-up.
Authors
- Gele B. Moloney (ORCID: https://orcid.org/0000-0003-2205-8994)
- Etk Lau
- Alan E. Wilson (ORCID: https://orcid.org/0000-0002-0079-6585)
Institutions
- University of Pittsburgh Medical Center (US)
Publication Details
- Journal
- JBJS Case Connector
- Published
- 2024-10-01
- DOI
- https://doi.org/10.2106/jbjs.cc.23.00523
- Primary Topic
- Hip and Femur Fractures
- Type
- article
- Field-Weighted Citation Impact
- 0.00