Evaluation and optimization of measurement conditions for heavy ions with fluorescent nuclear track detectors

In applying fluorescent nuclear track detectors for heavy ion measurement, the fluorescence intensity directly influences linear energy transfer (LET) estimation. However, the conventional analysis highly depends on the analyzed depth and the reference frames from the optical readout. To improve the accuracy of measurements of heavy ions, the gradient of intensity and moving average fluorescence intensity were acquired and calculated to optimize the measurement conditions. The results indicate that the intensity gradient is an alternative quantity to discriminate ions with different LET, and the moving average fluorescence intensity is readily differentiated among individual ions without overlapping values in fluorescence intensity. The scanned range of 25-45 μm is suggested to be the most suitable depth for measurement. To exclude the impact of the variation in fluorescence intensity on evaluation, it is recommended to use the quantities calculated in moving averages with depth in the future.

Authors

Institutions

Publication Details

Journal
Radiation Protection Dosimetry
Published
2024-11-01
DOI
https://doi.org/10.1093/rpd/ncae071
Citations
1
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
0.61

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation and optimization of measurement conditions for heavy ions with fluorescent nuclear track detectors

Hu Jun, Tamon Kusumoto, Satoshi Kodaira
1 citations
Radiation Protection Dosimetry
Radiation Detection and Scintillator Technologies
0.61
article

Evaluation and optimization of measurement conditions for heavy ions with fluorescent nuclear track detectors

Hu Jun, Tamon Kusumoto, Satoshi Kodaira
article en
1 citations

Abstract

In applying fluorescent nuclear track detectors for heavy ion measurement, the fluorescence intensity directly influences linear energy transfer (LET) estimation. However, the conventional analysis highly depends on the analyzed depth and the reference frames from the optical readout. To improve the accuracy of measurements of heavy ions, the gradient of intensity and moving average fluorescence intensity were acquired and calculated to optimize the measurement conditions. The results indicate that the intensity gradient is an alternative quantity to discriminate ions with different LET, and the moving average fluorescence intensity is readily differentiated among individual ions without overlapping values in fluorescence intensity. The scanned range of 25-45 μm is suggested to be the most suitable depth for measurement. To exclude the impact of the variation in fluorescence intensity on evaluation, it is recommended to use the quantities calculated in moving averages with depth in the future.

Radiation Protection DosimetryVol. 200(16-18)
National Institutes for Quantum Science and Technology (JP)
Japan Society for the Promotion of Science
Reduced inequalities
Openalex Percentile: Top 29%
Radiation Detection and Scintillator Technologies
0.61
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.